KNIFE, STRAIGHT
Received May 17, 2023 · Event occurred Sep 3, 2020
Report 9610617-2023-00621 · MDR key 16950067
Device
Generic name
Knife, Straight
Manufacturer
Karl Storz Se & Co. KgModel number
27068KCatalog number
27068K
Lot number
YV01
Product problems
- Fracture
- Fracture
Patient
Unknown
- Foreign Body In Patient
- Foreign Body In Patient
Narrative
Additional Manufacturer Narrative
BELATED EVALUATION AND REPORTING OF THIS COMPLAINT WAS DONE DURING RETROSPECTIVE REVIEW AS PART OF CAPA 20-0052 CORRECTIVE ACTION 6. THE SCALPEL WAS RETURNED WITH 2 PIECES. THE DISTAL KNIFE WAS BROKEN OFF FROM THE TUBE OF THE SCALPEL. UPON EVALUATION IT WAS FOUND THAT AT THE BROKEN AREA OF THE KNIFE THERE ARE SIGNS OF DISCOLORATION AND RUS. IT COULD BE THE EVIDENCE OF EITHER CORROSION OR EXERTING TOO MUCH FORCE TO THE WELDING JOINT. IN ADDITION, THE DISTAL END OF KNIFE WAS BEND WHICH COULD BE THE RESULT OF HIGH LATERAL FORCE APPLIED. THE ROOT CAUSE MOST PROBABLY WHICH MAY CAUSE DAMAGE AND BREAKAGE OF THE KNIFE IS DUE TO EXERTING TOO MUCH FORCE, ESPECIALLY LATERAL FORCE THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID (B)(4).
Description of Event or Problem
IT WAS REPORTED THAT THE SURGEON BECAME AWARE THAT THERE IS NO INSTRUMENT TIP AFTER THE SURGERY (OPTICAL URETHROTOMY). BROKEN TIP WAS FOUND IN PATIENTS BLADDER VIA X-RAY WHEN THE PATIENT WAS IN THE RECOVERY ROOM. LT WAS REMOVED BY CYSTOSCOPY WITH FOREIGN BODY REMOVAL AT ONCE. SECOND SURGERY (CYSTOSCOPY TO REMOVE REMAINING BROKEN TIP IN PATIENT BLADDER) WAS PERFORMED.