CADD INFUSION DISPOSABLES PORTFOLIO WITH NRFIT CON

Received May 11, 2023 · Event occurred Oct 25, 2021

Report 3012307300-2023-05525 · MDR key 16908907

Device

Generic name

Administrations Sets With Neuraxial Connectors

Manufacturer

St Paul

Model number

21-7605-24

Catalog number

21-7605-24

Product problems

  • Disconnection
  • Disconnection

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THIS REMEDIATION MDR WAS GENERATED UNDER PROTOCOL (B)(4), AS A RESULT OF WARNING LETTER CMS# 617147. ONE SAMPLE WAS DECONTAMINATED INSIDE A PLASTIC BAG WITHOUT ITS ORIGINAL PACKAGE. DURING THE VISUAL INSPECTION, THE SAMPLE RECEIVED DID NOT PRESENT ANY DAMAGE. DURING THE FUNCTIONAL TESTING NO PROBLEM WAS FOUND WITH THE ASSEMBLY OF THE CONNECTOR. THE COMPLAINT WAS NOT CONFIRMED. THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED AS COMPLAINT NOT BEING CONFIRMED. NO ACTION WAS TAKEN DUE TO COMPLAINT NOT CONFIRMED.

Description of Event or Problem

IT WAS REPORTED THAT THE THE DISPOSABLE CASSETTE TUBING BECAME DISCONNECTED FROM THE EXTENSIONS SET. NO PATIENT INJURY WAS REPORTED.