MalfunctionMSH

MASK PFR95 FS SMALL

Received May 8, 2023 · Event occurred Jan 14, 2021

Report 9616096-2023-00021 · MDR key 16885304

Device

Generic name

Icp Facial Protection Products

Manufacturer

O&m Halyard, Inc.

Model number

46827

Catalog number

991046827

Lot number

AM0178811

Product problems

  • Break
  • Break

Patient

Unknown

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

Narrative

Additional Manufacturer Narrative

TWO (2) SAMPLES ARRIVED WITH A HANDWRITTEN NUMBER WHICH MATCHES THE LOT NUMBER REPORTED. THESE SAMPLES APPEAR INTACT. THREE (3) SAMPLES ARRIVED WITH THE ELASTIC HEAD BAND COMPLETELY DETACHED FROM THE LEFT SIDE (AS WORN). FIVE (5) SAMPLES WERE INSPECTED AND APPEAR INTACT. THE ELASTIC HEAD BAND WAS TUGGED AS IF FOR DONNING. THERE WAS NO DETACHMENT OF THE ELASTIC HEAD BANDS. THE HEAD BANDS APPEAR SECURELY ATTACHED IN BOTH CORNER BONDS. THE DISPOSITION WILL BE CONFIRMED BECAUSE SOME SAMPLES ARRIVED WITH DETACHED HEAD BANDS. THERE IS NO DAMAGE OBSERVED ON THE BOND POINTS OF THE BLUE ELASTIC HEAD BANDS. THE DETACHMENT OF THE ELASTIC APPEARS TO COME FROM THE SEPARATION OF THE CORNER BOND AREAS. THE OPPOSITE CORNER BONDS WERE INSPECTED. THE CORNERS REMAINED BONDED AND THE ELASTIC HEAD BANDS APPEAR SECURELY ATTACHED. A REVIEW OF DHR PRODUCTION RECORDS WAS COMPLETED. PULL STRENGTH HEADSTRAP TESTS ARE CONDUCTED PER SPECIFICATION REQUIRING 4.0LBS PERFORMANCE STRENGTH. THE RECORDS FROM THIS LOT SHOW AVERAGE OF 5.0LBS (WITHIN SPECIFICATION). DOCUMENTED RECORDS SHOW ALL VISUAL AND FUNCTIONAL INSPECTIONS WERE WITHIN SPECIFICATIONS AND THERE WERE NO NONCONFORMANCES DOCUMENTED DURING PRODUCTION OF THE COMPLAINT LOT. NO ROOT CAUSE WAS IDENTIFIED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE O&M HALYARD, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT COMP (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21 CFR 803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BENCONSIDERED TO BE AN ADMISSION THAT A O&M HALYARD, INC. PRODUCT IS DEFECTIVE OR HAS CAUSED SERIOUS INJURY.

Description of Event or Problem

INCIDENT TWO. THE STRAPS WERE POPPING OFF AT THE SEAMS WHEN DONNING THE SMALL N95 MASK. IN LESS THAN A WEEK, WE ALSO HAD 2 INCIDENTS WHERE THE MASK BROKE WHILE IN A COVID PATIENTS ROOM. THE FOLLOWING MEDICAL DEVICE REPORT IS BEING FILED RETROACTIVE TO PRIOR DECISION THAT THIS MALFUNCTION WOULD BE UNLIKELY TO RESULT IN A SERIOUS INJURY OR ILLNESS. A REASSESSMENT OF THIS INCIDENT HAS SINCE FOUND THAT SHOULD THIS MALFUNCTION RECUR IN AN ISOLATION SETTING OR DURING DIRECT PATIENT CARE, THE END USER MAY CONTRACT COVID VIRUS AND THUS BE SUSCEPTIBLE TO A SERIOUS ILLNESS. FOR THIS REASON THIS INCIDENT IS BEING REPORTED AS A MALFUNCTION AT THIS TIME. PLEASE NOTE THAT IN THIS INCIDENT, THE REPORTER HAS INDICATED THAT NO INJURY OR ILLNESS RESULTED.