inforMED
MalfunctionMRZ

MEDFUSION 3000 SERIES

Received May 5, 2023 · Event occurred Sep 1, 2021

Report 3012307300-2023-05203 · MDR key 16878281

Device

Generic name

Accessories, Pump, Infusion

Manufacturer

St Paul

Catalog number

67-2515-51

Lot number

P0754077

Product problems

  • Mechanical Problem
  • Smoking
  • Mechanical Problem
  • Smoking

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

D4: UDI IS UNKNOWN. THIS REMEDIATION MDR WAS GENERATED UNDER PROTOCOL (B)(4), AS A RESULT OF WARNING LETTER CMS# (B)(4). A PRODUCT SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL AND FUNCTIONAL TESTING WERE PERFORMED. VISUALLY INSPECTED ALL BATTERIES, PLUGGED EACH BATTERY IN A TEST PUMP AND CHARGED FOR A FEW MINUTES, CHECKED BATTERY DIAGNOSTICS. DETECTED NO SMOKE OR ANY DAMAGE. ONLY 1 OUT OF 6 BATTERIES HAD "BATTERY NOT WORKING" ALARM OCCUR WITH ALL BATTERY VALUES ZERO. THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE CONFIRMED AS THE REPORTED ISSUE COULD NOT BE REPLICATED AT THE TIME OF THE DEVICE EVALUATION. ACTIONS WERE TAKEN TO MITIGATE THE REPORTED ISSUE: ALL 6 BATTERIES WERE SCRAPPED.

Description of Event or Problem

IT WAS REPORTED THAT THE BATTERIES CAUSED THE MACHINE SMOKE AND WOULDN'T WORK. NO PATIENT INJURY REPORTED.