VISTA CTO
Received May 2, 2023 · Event occurred Mar 2, 2023
Report 16852689 · MDR key 16852689
Device
Generic name
Orthosis, Cervical
Manufacturer
Aspen Medical Products, LlcModel number
984510Catalog number
984510
Product problems
- Structural Problem
Patient
NA · Unknown
- Insufficient Information
- Insufficient Information
Narrative
Description of Event or Problem
POSTERIOR HEAD WOUND WAS FOUND ABOVE CTO-BRACE ON OCCIPUT (BACK OF HEAD). WHEN THEY COULD START SITTING HIM UP WITH THE CTO DEVICE, THEY DID. HOWEVER, THE CTO DEVICE HAS A WAIST STRAP AND IF THE PATIENT HAS A LARGER ABDOMEN IT WILL PUSH UP ON THE WHOLE DEVICE INCLUDING ON THE MANDIBLE AND BACK OF THE HEAD. THE PADDING ON THE COLLAR IS NOT PROTECTIVE ENOUGH AND THE HARD PIECE CAN SNEAK OVER THE PADDING. HANGER CLINIC DID COME BACK TO ADJUST THE DEVICE AND PATIENT KEPT TRYING TO TAKE THE BRACE OFF BECAUSE IT WAS SO UNCOMFORTABLE. PHOTOS AVAILABLE.