inforMED
InjuryJDP

VA-LCP CONDYLAR PLATE 4.5/5.0 R 6HO L159

Received Apr 21, 2023 · Event occurred Jan 1, 2023

Report 8030965-2023-05084 · MDR key 16788153

Device

Generic name

Condylar Plate Fixation Implant

Manufacturer

Synthes Gmbh

Catalog number

02.124.406

Lot number

614P789

Product problems

  • Break
  • Break

Patient

Female

  • Nonunion/Delayed-union Bone Fracture
  • Nonunion/Delayed-union Bone Fracture

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT #(B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. ADDITIONAL NARRATIVE: ADDITIONAL PRO-CODES: HWC, HRS. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. REPORTER IS A J&J SALES REPRESENTATIVE. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN TAIWAN AS FOLLOWS: IT WAS REPORTED ON A UNKNOWN DATE THAT A PATIENT UNDERWENT SURGERY WITH VA-LCP CONDYLAR PLATE ON (B)(6) A NON-UNION AND PLATE BREAK WAS FOUND, THEREFOR, A REVISION SURGERY PROCEDURE WAS DONE ON (B)(6) 2023 TO CORRECT THE BREAKAGE. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) VA-LCP CONDYLAR PLATE 4.5/5.0 R 6HO L159 THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. ADDITIONAL NARRATIVE: DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

PLATE BREAKAGE WAS NOTED ON MARCH 31, 2023.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE DEVICE WAS BROKEN IN TWO PIECES. IT APPEARS THE BREAKAGE ORIGINATED FROM ONE OF THE LOCKING HOLES. NO OTHER PROBLEM IDENTIFIED. A DIMENSIONAL INSPECTION WAS NOT PERFORMED AS IT IS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCES. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE VA-LCP CONDYLAR PLATE 4.5/5.0 R 6HO L159, P/N: 02.124.406, LOT: 614P789, WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DRAWING/SPECIFICATIONS REVIEWED: YES. DIMENSIONAL INSPECTION: N/A. DEVICE HISTORY LOT: PRODUCT CODE: 02.124.406. LOT NUMBER: 614P789. MANUFACTURING SITE: MEZZOVICO. RELEASE TO WAREHOUSE DATE: 03 FEB 2022. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE NOT STERILE FINISHED LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.