inforMED
MalfunctionIOY

SUPPORT ARM 177

Received Apr 21, 2023 · Event occurred Mar 27, 2023

Report 8010042-2023-00768 · MDR key 16782864

Device

Generic name

Support, Arm

Model number

SUPPORT ARM 177

Catalog number

6481720

Product problems

  • Fracture
  • Fracture

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE SUPPORT ARM FOR PATIENT CIRCUIT BROKE. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF #: (B)(4).

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE SUPPORT ARM BRACKET BROKE. OUR FIELD SERVICE ENGINEER INVESTIGATED THE VENTILATOR ON SITE AND SOLVED THE ISSUE BY REPLACING THE SUPPORT ARM BRACKET. REGULAR FUNCTIONAL TESTS WERE CARRIED OUT AND THE FAILURE DID NOT CAUSE ANY INJURIES TO PATIENTS OR OPERATORS OR DELAYS IN PROCEDURES. A BROKEN SUPPORT ARM MAY LEAD TO TO STOP OF VENTILATION (EXTUBATION) OR INJURY. THE ROOT CAUSE AS TO WHY THE SUPPORT ARM BRACKET BROKE CANNOT BE ESTABLISHED BY THIS INVESTIGATION.

Description of Event or Problem

MANUFACTURERS REF: #(B)(4).