inforMED
MalfunctionMPO

NOVAPLUS INFANT HEEL WARMER SQUEEZE

Received Apr 14, 2023 · Event occurred Mar 21, 2023

Report 1423537-2023-00903 · MDR key 16747813

Device

Generic name

Infant Heel Warmer (chemical Heat Pack)

Manufacturer

Mo016 Mo-moberly

Model number

V11460-010

Catalog number

V11460-010

Lot number

V2S056

Product problems

  • Burst Container or Vessel
  • Burst Container or Vessel

Patient

Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

CUSTOMER REPORTED, THAT THE NICU TEAM HAD INFANT HEEL WARMERS EXPLODE WHILE BEING ACTIVATED. THE GEL LANDED ON THE NURSE¿S MASK AND FACE SHIELD. NO INJURY WAS REPORTED. THEY HAVE PULLED THE FOLLOWING LOT FROM ALL BABY UNITS.

Additional Manufacturer Narrative

A SAMPLE WAS RECEIVED FOR EVALUATION. AND THE INVESTIGATION DETERMINED, THAT THE SAMPLE RECEIVED WAS ALREADY ACTIVATED AND DEFECT REPORTED WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A BURST PACK, DUE TO AN INCOMPLETE TOP SEAL. THE MACHINE WAS SETUP CORRECTLY AND IS DESIGNED WITH A MECHANISM, WHICH REDUCES THE LIKELIHOOD FOR THIS ISSUE. DEVICE HISTORY RECORD REVIEW WAS COMPLETED ON THE REPORTED LOT V2S056. AND THE LOT WAS MANUFACTURED AND RELEASED IN COMPLIANCE WITH ALL REQUIREMENTS. NO ANOMALIES WERE FOUND DURING REVIEW OF THE RECORDS. CARDINAL HEALTH WILL CONTINUE TO MONITOR COMPLAINT TRENDS FOR THIS REPORTED ISSUE OF BURST AND WORK TO IDENTIFY IMPROVEMENT ACTIVITIES TO MINIMIZE REOCCURRENCE.