LUCEA 40
Received Apr 5, 2023 · Event occurred Mar 28, 2023
Report 9710055-2023-00287 · MDR key 16684234
Device
Product problems
- Fracture
- Detachment of Device or Device Component
- Fracture
- Detachment of Device or Device Component
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR EXAMINATION LIGHTS ¿ LUCEA 40. AS IT WAS STATED THE HEADLIGHT COVER WAS BROKEN RESULTING IN MISSING PARTICLES. GETINGE TECHNICIAN ALSO STATED THAT THE COVERS WERE BROKEN DUE TO THE USER NOT USING THE HANDLE ON THE LIGHT AND GRABBING THE LIGHT BY THE LENS. THERE WAS NO INJURY REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF DURING EXAMINATION MAY LEAD TO POTENTIAL INFECTION OF THE PATIENT. ACCORDING TO THE INFORMATION PROVIDED BY GETINGE TECHNICIAN, THE AFFECTED PARTS (ARD368605998 - HANDLE INTERFACE WITH FORK - LUCEA 40, ARD368606555 - TRANSPARENT PLASTIC COVER - LUCEA 40) HAVE BEEN REPLACED. IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE EXAMINATION LIGHT DID NOT MEET ITS SPECIFICATION DUE TO CRACKS AND MISSING PARTICLES FROM COVERS, WHICH CONTRIBUTED TO THE EVENT. PROVIDED INFORMATION DOES NOT INDICATE IF UPON THE EVENT OCCURRENCE THE DEVICE WAS OR WAS NOT BEING USED FOR PATIENT TREATMENT. REVIEW OF RECEIVED CUSTOMER PRODUCT COMPLAINTS RELATED TO INVESTIGATED ISSUE, REVEALED THAT THERE WERE NO INJURIES TO A USER NOR TO A PATIENT OR OPERATOR WHEN THIS PARTICULAR MALFUNCTION OCCURRED. WE CAN ASSUME THAT THE FAILURE RATIO OF CRACKS AND MISSING PARTICLES ON LUCEA 10/40 DEVICES IS MODERATE. MALFUNCTION INVESTIGATED HEREIN WAS FURTHER ANALYSED BY SUBJECT MATTER EXPERTS AT MAQUET SAS. AS THEY STATED, BASED ON SOME INTERNAL TEST DONE BY MAQUET SAS (E.G. REF: CRE 12-085), ONLY ABNORMAL USE (VIOLENT COLLISIONS, EXCESSIVE PRESSURE¿) OR THE USE OF INCOMPATIBLE CLEANING PRODUCTS OR PROTOCOL, CAN DAMAGE THIS DEVICE AS REPORTED IN THIS COMPLAINT. THE USER MANUAL EXPLAINS HOW TO CHECK THE LIGHT HEADS DURING DAILY INSPECTION. TO AVOID ANY SIMILAR INCIDENT, THE LUCEA PRODUCT MUST BE USED ACCORDING WITH THE USER MANUAL INFORMATION. WE BELIEVE THE RELATED DEVICES ARE PERFORMING CORRECTLY IN THE MARKET. WE ALSO BELIEVE THAT IF THE MANUFACTURER RECOMMENDATION HAD BEEN FOLLOWED THE INCIDENT COULD HAVE BEEN AVOIDED. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DO NOT PROPOSE ANY FURTHER ACTION AT THIS TIME. THE CORRECTION OF B5 DESCRIBE EVENT AND PROBLEM DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS B5 DESCRIBE EVENT AND PROBLEM: ON 28TH MARCH, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR EXAMINATION LIGHTS ¿ LUCEA 40. AS IT WAS STATED THE HEADLIGHT COVER WAS BROKEN RESULTING IN MISSING PARTICLES. THERE WAS NO INJURY REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF DURING EXAMINATION MAY LEAD TO POTENTIAL INFECTION OF THE PATIENT. CORRECTED B5 DESCRIBE EVENT AND PROBLEM: ON 21TH MARCH, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR EXAMINATION LIGHTS ¿ LUCEA 40. AS IT WAS STATED THE HEADLIGHT COVER WAS BROKEN RESULTING IN MISSING PARTICLES. THERE WAS NO INJURY REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF DURING EXAMINATION MAY LEAD TO POTENTIAL INFECTION OF THE PATIENT.
Description of Event or Problem
ON 21TH MARCH, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR EXAMINATION LIGHTS ¿ LUCEA 40. AS IT WAS STATED THE HEADLIGHT COVER WAS BROKEN RESULTING IN MISSING PARTICLES. THERE WAS NO INJURY REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF DURING EXAMINATION MAY LEAD TO POTENTIAL INFECTION OF THE PATIENT.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. DEVICE NOT RETURNED TO MANUFACTURER.
Description of Event or Problem
ON 28TH MARCH, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR EXAMINATION LIGHTS ¿ LUCEA 40. AS IT WAS STATED THE HEADLIGHT COVER WAS BROKEN RESULTING IN MISSING PARTICLES. THERE WAS NO INJURY REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF DURING EXAMINATION MAY LEAD TO POTENTIAL INFECTION OF THE PATIENT.