SUPPORT ARM 178
Received Apr 5, 2023 · Event occurred Mar 27, 2023
Report 8010042-2023-00695 · MDR key 16681671
Device
Generic name
Support, Arm
Manufacturer
Maquet Critical Care AbModel number
SUPPORT ARM 178Catalog number
6693715
Product problems
- Fracture
- Fracture
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
IT WAS CLAIMED THAT SUPPORT ARM 178 MOUNTED ON SERVO-U VENTILATOR HAS BEEN BROKEN. THERE WAS NO PATIENT HARM. BASED ON TECHNICIAN STATEMENT, THE THERE IS NO INFORMATION HOW BREAKAGE OCCURRED. ACCORDING TO THE PROVIDED PHOTO, SUPPORT ARM BROKE ON THE LOCKING JOINT. WHOLE ARM HAS BEEN REPLACED TO RESOLVE THE ISSUE. THE DEVICE WAS DELIVERED TO THE USER IN WORKING CONDITION. BROKEN/DAMAGED SUPPORT ARM MAY LEAD TO STOP OF VENTILATION (EXTUBATION) OR INJURY. THE PART HAS BEEN NOT AVAILABLE FOR THE FURTHER ANALYZE OF THE ISSUE. THEREFORE, THE ROOT CAUSE OF THE ISSUE COULD NOT BE DETERMINED.
Description of Event or Problem
IT WAS REPORTED THAT THE SUPPORT ARM HAS BEEN BROKEN. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF #: (B)(4).