inforMED
MalfunctionCDX

ELECSYS T4

Received Mar 31, 2023 · Event occurred Mar 3, 2023

Report 1823260-2023-01050 · MDR key 16658616

Device

Generic name

Radioimmunoassay, Total Thyroxine

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

07027885190

Lot number

65318702

Product problems

  • High Test Results
  • High Test Results

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

TWO PATIENT SAMPLES WERE PROVIDED FOR INVESTIGATION. AN INTERFERING SUBSTANCE TO THE RUTHENIUM COMPONENT OF THE REAGENT WAS DETECTED. PER PRODUCT LABELING, "IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN." THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative

THE SAMPLE WAS REQUESTED FOR A SAMPLE INTERFERENCE INVESTIGATION.THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS T4 ASSAY RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 6000 E601 MODULE AND A COBAS E 801 ANALYTICAL UNIT. THIS MEDWATCH WILL COVER T4. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6).FOR INFORMATION ON THE FT3 III RESULTS. THE T4 RESULT ON THE E601 ANALYZER WAS >320 NMOL/L. THE SAMPLE WAS TESTED ON ANOTHER LABORATORY'S E801 ANALYZER AND THE RESULT WAS 186 NMOL/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE E601 ANALYZER SERIAL NUMBER (B)(4). THE OTHER LABORATORY'S E801 ANALYZER SERIAL NUMBER WAS REQUESTED BUT NOT PROVIDED.