ELECSYS FT3 III
Received Mar 31, 2023 · Event occurred Mar 3, 2023
Report 1823260-2023-01039 · MDR key 16657587
Device
Generic name
Total Triiodothyronine
Manufacturer
Roche DiagnosticsModel number
NACatalog number
09005811190
Lot number
ASKU, 611920, 611918
Product problems
- High Test Results
- High Test Results
Patient
87 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE COMPONENT, EVALUATION METHOD, EVALUATION RESULT, AND EVALUATION CONCLUSION CODES HAVE BEEN UPDATED.
Additional Manufacturer Narrative
THE PATIENT SAMPLE WAS FURTHER INVESTIGATED AND IT WAS DETERMINED THE SAMPLE CONTAINED AN INTERFERING FACTOR AGAINST THE FT3 ASSAY ANTIBODY/IDIOTYPE. PER PRODUCT LABELING: "IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN." THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ELECSYS FT3 III VER. 2 (FT3 III V 2) ON TWO COBAS E 801 MODULES AND A COBAS E 411 IMMUNOASSAY ANALYZER. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. REFER TO THE ATTACHMENT FOR ALL PATIENT DATA. QUESTIONED VALUES ARE HIGHLIGHTED IN YELLOW. THE SAMPLE WAS INITIALLY TESTED WITH THE FT3 II V 2 ASSAY WHEN TESTED ON THE CUSTOMER'S E 801 MODULE ON (B)(6) 2023. THE SAMPLE WAS REPEATED BY THE CUSTOMER USING THE WAKO ACCURASEED FT3 METHOD. THE CUSTOMER ALSO TREATED THE SAMPLE WITH 25 % POLYETHYLENE GLYCOL (PEG) AND REPEATED IT ON THEIR E 801 MODULE. THE SAMPLE WAS PROVIDED FOR INVESTIGATION WHERE IT WAS TESTED WITH FT3 III V 2 ON A SECOND E 801 ANALYZER AND AN E411 ANALYZER ON (B)(6) 2023. THE SAMPLE WAS ALSO REPEATED USING THE ABBOTT ARCHITECT FT3 METHOD ON (B)(6) 2023. THE SERIAL NUMBER OF THE CUSTOMER'S E 801 MODULE WAS REQUESTED, BUT NOT PROVIDED. THE FT3 III V 2 REAGENT LOT NUMBER AND EXPIRATION DATE USED ON THIS ANALYZER WERE REQUESTED, BUT NOT PROVIDED. THE SERIAL NUMBER OF THE E 801 MODULE USED FOR INVESTIGATION IS (B)(4). FT3 III V 2 REAGENT LOT NUMBER 611920, WITH AN EXPIRATION DATE OF 31-MAY-2023 WAS USED ON THIS ANALYZER. THE SERIAL NUMBER OF THE E411 ANALYZER USED FOR INVESTIGATION IS (B)(4). FT3 III V 2 REAGENT LOT NUMBER 611918, WITH AN EXPIRATION DATE OF 30-MAY-2023 WAS USED ON THIS ANALYZER.
Additional Manufacturer Narrative
THE PATIENT SAMPLE WAS REQUESTED FOR INVESTIGATION.