CLIN CHEM ALBUMIN BCG REAGENT

Received Apr 12, 2010 · Event occurred Feb 18, 2010

Report 2018433-2010-00064 · MDR key 1656698

Device

Generic name

Quantitation Of Albumin In Human Serum Or Plasma

Catalog number

7D53-21

Product problems

  • Output above Specifications

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE CUSTOMER STATED A SHIFT OUT OF RANGE HIGH FOR THE CLINICAL CHEMISTRY ALBUMIN BCG QUALITY CONTROLS (QC) AFTER CALIBRATION. THE CUSTOMER HAD BEEN RUNNING THE ALBUMIN ASSAY ALL MORNING WITHOUT ERROR, THEN, THE CALIBRATION EXPIRED. RECALIBRATION OF THE ASSAY PASSED, HOWEVER, THE QC WAS VERY HIGH OUT OF RANGE WITH NO CHANGE TO THE CALIBRATOR LOT OR REAGENT KITS. THE CUSTOMER VERIFIED THE CORRECT CALIBRATOR SET POINTS AND PERFORMED OTHER RECOMMENDED TROUBLESHOOTING. THE CUSTOMER STATED RECALIBRATION PASSED SPECIFICATIONS WHEN THE CORRECT CALIBRATOR LEVEL WAS USED. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Additional Manufacturer Narrative

(B)(4). CONCOMITANT MEDICAL PRODUCTS: ARCHITECT C8000 ANALYZER, LIST # 1G06-01, (B)(4). THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Remedial action

  • Recall