SUPPORT ARM 177
Received Mar 15, 2023 · Event occurred Mar 10, 2023
Report 8010042-2023-00561 · MDR key 16545568
Device
Generic name
Support, Arm
Manufacturer
Maquet Critical Care AbModel number
SUPPORT ARM 177Catalog number
6481720
Product problems
- Fracture
- Fracture
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
IT WAS REPORTED THAT THE SUPPORT ARM FOR PATIENT TUBES BROKE. NO REPAIR WAS REQUESTED AND NO PARTS WERE RETURNED. THE SUPPORT ARM IS A CASTING AND IT MOST PROBABLY DEVELOPED A CRACK AT AN EARLIER OCCASION EITHER BY OVERLOADING OR IMPACT AND THIS LED TO THE REPORTED BREAKING. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGNED ACCORDING TO STANDARD. PREVIOUS INVESTIGATIONS OF SIMILAR FAULTS LED TO A REDESIGN AND CHANGE OF THE MANUFACTURING PROCESS IN ORDER TO OBTAIN A HIGHER MECHANICAL STRENGTH OF THE SUPPORT ARM. THE SUPPORT ARM ASSOCIATED WITH THIS COMPLAINT HAD BEEN MANUFACTURED BEFORE THIS CHANGE WAS IMPLEMENTED IN PRODUCTION. TYPE OF INVESTIGATION: 4117.
Description of Event or Problem
IT WAS REPORTED THAT THE SUPPORT ARM BROKE. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF #: (B)(4).