inforMED

LINEAR ACCELERATOR TANGENTIAL BREAST BEAM COLLIMATOR

Received Oct 3, 1994

Report MW1003570 · MDR key 16544

Device

Model number

597-014

Product problems

  • Inadequacy of Device Shape and/or Size

Patient

*

Narrative

Description of Event or Problem

IT HAS BEEN NOTED THAT THE BLOCK IS NOT LONG ENOUGH TO COMPLETELY BLOCK THE RADIATION FIELD WHEN TREATMENT IS PERFORMED WITH THE COLLIMATER WIDE OPEN, THEREFORE, A SMALL AMOUNT OF FLASH RADIATION HITS THE PT IN AN UNINTENDED AREA. THIS SITUATION ONLY OCCURS WITH TANGENTIAL BREAST SETUPS.