OLYMPUS
Received Apr 27, 1998
Report 9610773-1998-00009 · MDR key 165241
Device
Generic name
Obturator, 3 Devices
Manufacturer
Olympus Winter & Ibe GmbhModel number
A2637Catalog number
83A2637
Lot number
76W, 1 DEVICE
Patient
UNKNOWN
Narrative
Additional Manufacturer Narrative
UPON EVAL, THE MFR FOUND THAT THE EPOXY CEMENT USED TO SECURE AN INTERNAL SCREW IN THE DEFLECTING TIP OF THE OBTURATOR HAD DETERIORATED OVER TIME. THIS ALLOWED THE SCREW TO SHIFT, AND AS A RESULT, THE OBTURATOR LOCKED IN A DEFLECTED POSITION. TO COUNTER THIS PHENOMENON, THE MFR HAS VALIDATED A CHANGE TO A LASER WELDING PROCESS.
Description of Event or Problem
THE CUSTOMER RETURNED AN A2637 OBTURATOR BECAUSE IT WAS FROZEN IN POSITION AND WOULD NOT ANGULATE (DEVICE 1). OLYMPUS CONTACTED THE CUSTOMER TO OBTAIN ADD'L INFO REGARDING THE DEVICE. DEVICES LOCKED UP DURING PREPARATION BEFORE A MEDICAL PROCEDURE AND WAS NOT USED IN THE PROCEDURE.