ELECSYS FT3 III
Received Mar 8, 2023 · Event occurred Jan 28, 2023
Report 1823260-2023-00723 · MDR key 16501015
Device
Generic name
Total Triiodothyronine
Manufacturer
Roche DiagnosticsModel number
NACatalog number
09005811190
Lot number
611920, 611918
Product problems
- High Test Results
- High Test Results
Patient
57 YR · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
FURTHER INVESTIGATIONS OF THE PATIENT SAMPLE DETERMINED THAT IT DOES NOT CONTAIN AN INTERFERING FACTOR AGAINST A COMPONENT OF THE ROCHE ASSAY. THE VALUE DIFFERENCES OBTAINED WITH THE DIFFERENT ASSAYS RELATE TO THE DIFFERENCES IN THE SETUP OF THE ASSAYS, THE ANTIBODIES USED, AND DIFFERENCES IN THE STANDARDIZATION MATERIALS AND PROCEDURES USED. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ELECSYS FT3 III VER. 2 (FT3 III V 2) ON TWO COBAS E 801 MODULES AND A COBAS E 411 IMMUNOASSAY ANALYZER. THE SAMPLE ALSO HAD DISCREPANT RESULTS FOR ELECSYS FT3 III (FT3 III) WHEN TESTED ON THE SECOND E 801 MODULE AND THE E 411 ANALYZER THAT WERE USED FOR INVESTIGATION. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THIS MEDWATCH WILL APPLY TO THE FT3 III V 2 ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE FT3 III ASSAY. REFER TO THE ATTACHMENT FOR ALL PATIENT DATA. QUESTIONED VALUES ARE HIGHLIGHTED IN YELLOW. THE SAMPLE WAS INITIALLY TESTED WITH THE FT3 II V 2 ASSAY WHEN TESTED ON THE CUSTOMER'S E 801 MODULE ON (B)(6) 2023. THE SAMPLE WAS REPEATED BY THE CUSTOMER USING THE WAKO ACCURASEED FT3 METHOD. THE CUSTOMER ALSO TREATED THE SAMPLE WITH POLYETHYLENE GLYCOL (PEG) AND REPEATED IT ON THEIR E 801 MODULE. THE SAMPLE WAS PROVIDED FOR INVESTIGATION WHERE IT WAS TESTED WITH FT3 III AND FT3 III V 2 ON A SECOND E 801 ANALYZER AND AN E411 ANALYZER ON (B)(6) 2023. THE SAMPLE WAS ALSO REPEATED USING THE ABBOTT ARCHITECT FT3 METHOD ON (B)(6) 2023. THE SERIAL NUMBER OF THE CUSTOMER'S E 801 MODULE IS (B)(4). THE FT3 III V2 REAGENT LOT NUMBER AND EXPIRATION DATE USED ON THIS ANALYZER WERE NOT PROVIDED. THE SERIAL NUMBER OF THE E 801 MODULE USED FOR INVESTIGATION IS (B)(4). FT3 III V 2 REAGENT LOT NUMBER 611920, WITH AN EXPIRATION DATE OF 31-MAY-2023 WAS USED ON THIS ANALYZER. THE SERIAL NUMBER OF THE E411 ANALYZER USED FOR INVESTIGATION IS (B)(4). FT3 III V 2 REAGENT LOT NUMBER 611918, WITH AN EXPIRATION DATE OF 30-MAY-2023 WAS USED ON THIS ANALYZER.