OLYMPUS
Received Apr 24, 1998 · Event occurred Mar 4, 1998
Report 9610773-1998-00008 · MDR key 164784
Device
Generic name
Obturator
Manufacturer
Olympus Winter & Ibe GmbhModel number
A2636Catalog number
83A2636
Lot number
NA
Patient
UNKNOWN
Narrative
Additional Manufacturer Narrative
UPON EVAL, THE MFR FOUND THAT THE EPOXY CEMENT USED TO SECURE AN INTERNAL SCREW IN THE DEFLECTING TIP OF THE OBTURATOR HAD DETERIORATED OVER TIME. THIS ALLOWED THE SCREW TO SHIFT, AND AS A RESULT, THE OBTURATOR LOCKED IN A DEFLECTED POSITION. TO COUNTER THIS PHENOMENON, THE MFR HAS VALIDATED A CHANGE TO A LASER WELDING PROCESS.
Description of Event or Problem
THE CUSTOMER RETURNED AN OBTURATOR TO CO'S REPAIR FACILITY, BECAUSE THE TIP WOULD NOT ANGULATE. THE OBTURATOR WAS LOCKED IN POSITION. ON MARCH 4, 1998 THE OBTURATOR LOCKED IN POSITION DURING PREPARATION, PRIOR TO THE PROCEDURE TO BE PERFORMED. A SECOND DEVICE WAS USED TO PERFORM THE PROCEDURE AND NO PROBLEMS WERE ENCOUNTERED. THE CUSTOMER WOULD NOT RELEASE ANY INFO REGARDING THE PROCEDURE BEING PERFORMED, SINCE THE FAILED DEVICE WAS NOT USED DURING THE PROCEDURE.