YOMI ROBOTIC ASSISTANT

Received Feb 28, 2023 · Event occurred Oct 11, 2022

Report MW5115342 · MDR key 16462002

Device

Generic name

Dental Navigation System

Manufacturer

Neocis Inc.

Model number

DRW-1000-00

Product problems

  • Product Quality Problem
  • Calibration Problem

Patient

57 YR · Male

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

AFTER USING THE YOMI ROBOTIC-ASSISTED IMPLANT PLACEMENT DEVICE TO PLACE AN IMPLANT IN THE SITE OF TOOTH NUMBER 19, I NOTED IN THE POSTOPERATIVE RADIOGRAPH THAT IMPLANT ANGULATION WAS CORRECT, BUT THAT THE IMPLANT WAS MORE 2 MM ANTERIOR TO THE PLANNED POSITION. THIS WILL NOT PRESENT A PROBLEM WITH RESTORING THIS PARTICULAR IMPLANT WITH A CROWN, BUT THE NOTABLE ANTERIOR-POSTERIOR DEVIATION WOULD BE A PROBLEM IN SITES THAT REQUIRE A MORE PRECISE POSITION AND RELY ON THE NEOCIS CLAIM OF SUBMILLIMETRIC ACCURACY. PRIOR TO DISENGAGING THE ROBOT'S TRACKER ARM, I CAREFULLY CHECKED THE ATTACHMENT OF THE TOOTH LINK MECHANISM TO THE OPPOSITE SIDE OF THE DENTAL ARCH AND IT WAS SECURE, WHICH RULES OUT A LOOSE CONNECTION. THE ROBOT WAS PROPERLY CALIBRATED WHEN NEOCIS CHECKED IT. I AND MY STAFF HAD RECENTLY BEEN FULLY TRAINED WITH USE OF THE ROBOT, AND WE WERE CAREFUL TO OBSERVE ALL PROPER PROTOCOLS. WE WERE REPEATEDLY PROMISED "SUBMILLIMETRIC ACCURACY" FOR DENTAL IMPLANT OSTEOTOMIES PREPARED WITH THIS ROBOTIC ASSISTANCE. THIS CASE NOT ONLY SHOWED FAILURE TO ACHIEVE SUBMILLIMETRIC ACCURACY, BUT IT WAS A DEVIATION THAT COULD HAVE ENDANGERED THE PATIENT WITH AN IMPLANT NOT IN THE CORRECT POSITION FOR RESTORING. I AM LEFT NOT KNOWING WHETHER THIS 2 MM DEVIATION WAS DUE TO A DEFICIENCY OF THE TECHNOLOGY ITSELF, A DEFECTIVE UNIT, IMPROPER CALIBRATION, MISUSE OF THE UNIT BY ME OR MY STAFF, IMPROPER TRAINING, OR A COMBINATION OF THE ABOVE. THIS ADVERSE EVENT WAS ACCENTUATED BY ADVERSE EVENTS WITH THE YOMI SURGICAL ROBOT ON NUMEROUS OTHER PATIENTS, WHICH I WILL DESCRIBE IN SEPARATE REPORTS. AFTER USING THE ROBOT FOR APPROXIMATELY 40 CASES OVER MORE THAN TWO MONTHS, I CONCLUDED THAT PATIENTS WERE BEING PUT AT RISK BY CONTINUED USE OF THE ROBOT. I INFORMED NEOCIS THAT I HAD STOPPED USING THE ROBOT BECAUSE I WAS CONCERNED ABOUT PATIENT SAFETY. NEOCIS BEGRUDGINGLY ACCEPTED THE UNIT BACK, BUT THEIR ATTITUDE SEEMED ONE OF INDIFFERENCE ABOUT MY SAFETY CONCERNS, SEEMINGLY BECAUSE SALES ARE GOOD, THEY "CAN'T MAKE THEM FAST ENOUGH", AND "THIS JUST DOESN'T HAPPEN". I AM CONCERNED ABOUT PATIENT SAFETY WITH THE YOMI ROBOT IN CASES SIMILAR TO THE ONES I WAS DOING, WHICH WERE PRIMARILY SINGLE IMPLANTS, WITH AN OCCASIONAL TWO- OR THREE-IMPLANT CASE, USING THE NEOCIS TOOTH LINK SYSTEM FOR ROBOT POSITIONING AND CALIBRATION. I AM ALSO CONCERNED ABOUT HOW NEOCIS PRESENTED FDA CONTRAINDICATIONS FOR USING THE YOMI DEVICE, WHICH INCLUDE PREEXISTING CROWNS, BRIDGES OR IMPLANTS ON THE OPPOSITE SIDE OF THE ARCH, OR TEMPOROMANDIBULAR JOINT PROBLEMS. A VERY SMALL PERCENTAGE OF POTENTIAL IMPLANT PATIENTS ARE WITHOUT ANY OF THESE CONTRAINDICATIONS, WELL UNDER FIVE PERCENT IN MY EXPERIENCE, SO THE OVERWHELMING MAJORITY OF PATIENTS TREATED NEED THE ROBOT USED IN AN OFF-LABEL CAPACITY. NEOCIS WAS NOT FORTHCOMING WITH THESE CONTRAINDICATIONS PRIOR TO DELIVERY OF THE UNIT, AND I DID NOT SEE ANY MENTION OF THEM IN MATERIAL I HAD ACCESS TO PRIOR TO PURCHASE AND DELIVERY OF THE ROBOT. BASED ON MY COLLECTIVE EXPERIENCE WITH THIS TECHNOLOGY, USING IT IN THE WAY WE WERE TRAINED AND CLOSELY FOLLOWING ALL PROTOCOLS, I CONCLUDED THAT THIS PRODUCT SHOULD PROBABLY NOT BE ON THE MARKET FOR SIMILAR DENTAL IMPLANT USE. NEOCIS CONTINUES TO MAKE MARKETING CLAIMS THAT WERE NOT SUBSTANTIATED CONSISTENTLY DURING OUR CLINICAL USE OF THE ROBOT. WE ENCOUNTERED TRUE AND ACCEPTABLE PRECISION IN ONLY 50-60% OF OUR YOMI IMPLANT CASES.