Received Mar 1, 2023 · Event occurred Sep 9, 2022

Report 1062191-2022-00001 · MDR key 16460357

Device

Generic name

Conserver, Oxygen

Manufacturer

Inovo, Inc

Model number

OM-900A

Product problems

  • Flare or Flash
  • Flare or Flash

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

INOVO INC. WAS NOTIFIED OF AN INCIDENT INVOLVING AN OXYGEN CONSERVER BY A PROVIDER WHO STATED THAT "A PATIENT HAD FITTED THE CONSERVER ONTO A CH MEDICAL OXYGEN CYLINDER BUT DID NOT TURN IT ON. APPROXIMATELY 24 HOURS LATER, THE PATIENT NEEDED TO VISIT THE GP AND WHEN OUTSIDE THE GP'S CLINIC THEY TURNED IT ON. AT THAT POINT THE PATIENT HEARD A HISSING NOISE AND THEN A BANG." THERE WAS NO REPORT OR EVIDENCE OF ILLNESS, INJURY OR MEDICAL TREATMENT ASSOCIATED WITH THE COMPLAINT. INOVO EVALUATED THE PRODUCT, WHICH REVEALED THAT THE IGNITION MECHANISM WAS FLOW FRICTION, WHICH CAN OCCUR IF THE SEAL WASHER WEARS OVER TIME, OR IF THE DEVICE IS NOT CONNECTED TO THE COMPRESSED OXYGEN CYLINDER TIGHTLY ENOUGH, BOTH OF WHICH ARE ADDRESSED BY THE DEVICE'S INSTRUCTIONS FOR USE.

Remedial action

  • Inspection