SOLAR GI HRM H2O 24P TROLLEY SYSTEM
Received Feb 27, 2023 · Event occurred Sep 15, 2022
Report 3017448360-2023-00012 · MDR key 16448312
Device
Generic name
System, Gastrointestinal Motility (electrical)
Manufacturer
Medical Measurement Systems B.v.Model number
G3-8Product problems
- Excess Flow or Over-Infusion
- Excess Flow or Over-Infusion
Patient
6 YR · Female
- Convulsion/Seizure
- Convulsion/Seizure
Narrative
Additional Manufacturer Narrative
THE PART WAS RECEIVED AT OUR FACILITIES AND SENT TO OUR CONTRACTED MANUFACTURER: DEMCON. THE INVESTIGATION DESCRIBED BELOW WAS CARRIED OUT BY DEMCON: REPAIR// MPP PLUS // SN: (B)(6); (MAIN UNIT: (B)96)) // PUMP HAS NEGATIVE PRESSURE DESCRIPTION COMPLAINT CUSTOMER : PUMP HAS NEGATIVE PRESSURE INVESTIGATION: THE PRESSURE FOR THE WATER CONTAINER IS SHOWING -490 MBAR. THERE IS DIRT IN THE TUBING TO THE PRESSURE SENSOR, SEE PICTURE (ATTACHED). CANNOT SEE IF THE VALVES ARE AFFECTED BY LIQUID. ZEROED THE PRESSURE SENSOR AND THE OUTGOING PRESSURE IS NOW EQUAL TO THE DISPLAYED PRESSURE. POSSIBLE CAUSE OF THE PROBLEM: DUST, LIQUID. PRODUCTION YEAR: 2021-11-17. DESCRIPTION REPAIR: HAVE TO REPLACE A FEW TUBES IN THE UNIT. THEREFORE, WE CONCLUDE THAT THE ERROR WAS CAUSED BY LIQUID OR DIRT INGRESS INTO THE PRESSURE SENSOR TUBING.
Additional Manufacturer Narrative
DEVICE WILL BE RETURNED TO MANUFACTURER FOR EVALUATION BUT HAS NOT YET BEEN RECEIVED. AN INVESTIGATION IS PENDING.
Description of Event or Problem
THERE WAS A PATIENT IMPACT (WATER INTOXICATION & SEIZURE) AND THERE IS A FORMAL REVIEW PROCESS BEING CONDUCTED BY THE HOSPITAL. STUDY COMMENCED AT 13:40 ON (B)(6) 2022. THEY HAD A 6 YEAR-OLD FEMALE PATIENT WHO HAD A DROP IN THEIR SODIUM LEVELS FROM 138 MMOL/L (MEASURED PRIOR TO STUDY) TO 125MMOL/L (APPROX. 08:00 THE FOLLOWING MORNING). ACCORDING TO THE REPORT FROM THE DOCTOR, THE PATIENT DEVELOPED A SEIZURE FROM IATROGENIC HYPONATREMIA - WATER INTOXICATION. THE REMAINDER OF THE TEST WAS THEN ABORTED IMMEDIATELY AFTERWARDS (08:45 THE FOLLOWING MORNING). OUR CUSTOMER HAS INDICATED THAT THERE WERE NO LONG TERM EFFECTS TO THE PATIENT THAT THEY ARE AWARE OF, HOWEVER THE INCIDENT IS UNDER REVIEW IN THE HOSPITAL AND A FORMAL INCIDENT REPORT IS BEING PREPARED.
Remedial action
- Repair