ELECSYS VITAMIN B12 II
Received Feb 25, 2023 · Event occurred Jan 30, 2023
Report 1823260-2023-00588 · MDR key 16440916
Device
Generic name
Radioassay, Vitamin B12
Manufacturer
Roche DiagnosticsModel number
B12 G2Catalog number
07212771190
Lot number
66190705
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- High Test Results
- High Test Results
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
CALIBRATION RESULTS FROM 26-JAN-2023 WERE BELOW THE EXPECTED RANGE FOR BOTH CALIBRATOR LEVELS. QUALITY CONTROL RESULTS WERE WITHIN THE EXPECTED RANGE. THERE IS NO INDICATION FOR A REAGENT ISSUE. THE LAST INSTRUMENT MAINTENANCE WAS PERFORMED AT THE BEGINNING OF JANUARY-2023. THE PATIENT¿S SAMPLE WAS REQUESTED BUT NOT AVAILABLE FOR INVESTIGATION. BOTH ROCHE AND COMPETITOR-GENERATED RESULTS MATCH IN THE RELATION TO THE RESPECTIVE EXPECTED RANGES. THE MEASURED VITAMIN B 12 VALUE OF A PATIENT¿S SAMPLE CAN VARY DEPENDING ON THE TESTING PROCEDURE USED. ASSAYS FROM DIFFERENT VENDORS CAN GENERATE DIFFERENT VALUES. THIS RELATES TO THE OVERALL SETUPS OF THE ASSAYS, THE ANTIBODIES USED AND, DIFFERENCES IN REFERENCE MATERIALS/METHODS, AND THE STANDARDIZATION METHODOLOGY USED. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
NA.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS VITAMIN B12 II ASSAY ON A COBAS E 411 ANALYZER (SERIAL NUMBER UNKNOWN) AND COMPARED TO A COMPETITOR METHOD. THE SAMPLE INITIALLY RESULTED IN A VITAMIN B 12 II VALUE OF 640 PG/ML ON THE COBAS E 411 ANALYZER. THE REPORTER MENTIONED THAT THIS RESULT WAS REPRODUCIBLE. NO SPECIFIC VALUE WAS PROVIDED. THE SAMPLE WAS REPEATED WITH A COMPETITOR METHOD, RESULTING IN A VITAMIN B 12 VALUE OF 468 PG/ML. IT WAS ASKED BUT IT IS UNKNOWN WHICH RESULT IS CONSIDERED TO BE CORRECT. NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY.