MalfunctionMRZ
ELITE BIOMEDICAL BEZEL ASSEMBLY PUMP
Received Feb 23, 2023
Report MW5115145 · MDR key 16436598
Device
Generic name
Accessories, Pump, Infusion
Manufacturer
Elite Biomedical Solutions Llc.Product problems
- Excess Flow or Over-Infusion
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT MORPHINE INFUSED OVER FIVE (5) HOURS, HOWEVER WAS ORDERED TO INFUSE OVER TWENTY-FOUR (24) HOURS. THERE WAS NO PATIENT HARM. BD PERFORMED INVESTIGATION, IT WAS FOUND THE ROOT CAUSE OF THE REPORTED OVER INFUSION WAS DETERMINED TO BE DUE TO A THIRD-PARTY BEZEL ASSEMBLY. THE THIRD PART WAS MANUFACTURED BY ELITE BIOMEDICAL. THIRD PARTY PARTS HAVE NOT BEEN VALIDATED BY BD FOR SAFETY AND EFFICACY WITH ALARIS SYSTEM PRODUCTS.