BD BBL¿ ACRIDINE ORANGE STAIN
Received Feb 24, 2023 · Event occurred Jan 23, 2023
Report 1119779-2023-00175 · MDR key 16434405
Device
Generic name
Acridine Orange
Manufacturer
Becton, Dickinson & Co. (sparks)Model number
212537Catalog number
212537
Lot number
UNKNOWN
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H.6 INVESTIGATION SUMMARY: THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR COMPLAINT RELATED TO BOTTLE ACRIDINE ORANGE STAIN 250ML 4 EA, CATALOG NUMBER 212537, AND COMPLAINT #(B)(6) FOR ISSUES DIFFERENTIATING BACTERIA. COMPONENTS ARE MILLED AND BLENDED (WHERE REQUIRED) AND DISPENSED INTO CONTAINERS. FOLLOWING QC RELEASE, AND BASED UPON INVENTORY DEMAND, FINAL PACKAGING OPERATIONS OCCUR, WHICH INCLUDE DISPENSING INTO AND LABELING OF FINAL CONFIGURATIONS, FOLLOWED BY TRANSPORT TO THE DISTRIBUTION CENTER. COMPLAINT INVESTIGATION TYPICALLY INVOLVES A REVIEW OF THE COMPLAINT BATCH HISTORY RECORD. NO COMPLAINT LOT OF MATERIAL 212537 WAS PROVIDED, THEREFORE NO BATCH HISTORY REVIEW COULD BE PERFORMED. RELEASE TESTING FOR THIS PRODUCT INCLUDES SOLUTION APPEARANCE, AND CULTURAL RESPONSE WITH ORGANISMS AS SPECIFIED ON THE BD CERTIFICATE OF ANALYSIS. ALL QC RESULTS MUST MEET THE COA SPECIFICATIONS PRIOR TO RELEASE. COMPLAINT TRENDS WERE REVIEWED FOR A PERIOD COVERING 12 MONTHS. DURING THAT TIME, THERE HAVE BEEN NO COMPLAINTS ON THIS PRODUCT FOR ANY REASON. NO PHOTOS OR RETURNS WERE RECEIVED. THIS COMPLAINT IS NOT CONFIRMED. H3 OTHER TEXT : SEE H.10.
Description of Event or Problem
IT WAS REPORTED WHEN USING BD BBL¿ ACRIDINE ORANGE STAIN HAS PERFORMANCE ISSUES. THE FOLLOWING INFORMATION WAS PROVIDE THE INTIAL REPORTER: RAPID TEST FOR PRESENCE OF MICROORGANISMS (BACTERIA, YEAST, MOLD) LIMITATIONS OF METHOD MULTIPLE STUDIES MENTIONED ACRIDINE ORANGE¿S INABILITY TO DISCRIMINATE BETWEEN A LIVE AND DEAD CELL. SAW POOR CORRELATION BETWEEN THIS METHOD AND INGREDIENTS WHERE THERE WAS A HIGH LEVEL OF NON-VIABLE BACTERIA DUE TO A PROCESS STEP (E.G. BLANCHING OF VEGETABLES).
Additional Manufacturer Narrative
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
Description of Event or Problem
IT WAS REPORTED WHEN USING BD BBL¿ ACRIDINE ORANGE STAIN HAS PERFORMANCE ISSUES THE FOLLOWING INFORMATION WAS PROVIDE THE INITIAL REPORTER: RAPID TEST FOR PRESENCE OF MICROORGANISMS (BACTERIA, YEAST, MOLD) LIMITATIONS OF METHOD MULTIPLE STUDIES MENTIONED ACRIDINE ORANGE¿S INABILITY TO DISCRIMINATE BETWEEN A LIVE AND DEAD CELL. SAW POOR CORRELATION BETWEEN THIS METHOD AND INGREDIENTS WHERE THERE WAS A HIGH LEVEL OF NON-VIABLE BACTERIA DUE TO A PROCESS STEP (E.G. BLANCHING OF VEGETABLES).