SENSICA UO SYSTEM
Received Feb 21, 2023 · Event occurred Jan 31, 2023
Report 1018233-2023-01019 · MDR key 16408581
Device
Generic name
Sensica Device
Manufacturer
C.r. Bard, Inc. (covington) -1018233Model number
SCCS1002Catalog number
SCCS1002
Lot number
NA
Product problems
- Failure to Calibrate
- Failure to Calibrate
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE REPORTED ISSUE WAS CONFIRMED. THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. A POTENTIAL ROOT CAUSE OF THE REPORTED ISSUE COULD BE DUE TO THE PART GEOMETRY / CIRCUIT DESIGN- TRACE MATERIAL SELECTION AND COMPONENT TOLERANCES. HOWEVER, THIS CANNOT BE CONFIRMED. THE DEVICE WAS EVALUATED UPON RECEIPT. THE SCREEN PROTECTOR NEEDS TO BE REPLACED. REPLACED SCREEN PROTECTOR. ADDITIONAL MAINTENANCE WAS PERFORMED. ADDED GASKET FOAM. COMPLETED FINAL FUNCTIONAL TESTING. THE DEVICE DID NOT MEET SPECIFICATIONS, AND WAS INFLUENCED BY THE REPORTED FAILURE. THE DEVICE WAS NOT IN USE ON A PATIENT. A REVIEW OF THE DEVICE HISTORY RECORDS DID NOT SHOW ANY PROBLEMS OR CONDITIONS THAT WOULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "THE IFU CURRENTLY INSTRUCTS THE USER ON THE PROPER METHOD TO USE THIS DEVICE TO AVOID UNDUE INJURY TO THE PATIENT AND DAMAGE TO THE PRODUCT. INDICATIONS FOR USE: INDICATIONS: THE BD SENSICA¿ URINE OUTPUT SYSTEM IS AN AUTOMATED SYSTEM FOR CONTINUOUS MONITORING OF URINE OUTPUT (UO) AND CORE BLADDER TEMPERATURE WHEN CONNECTED TO A TEMPERATURE-SENSING FOLEY CATHETER. TARGET POPULATION: THE BD SENSICA¿ URINE OUTPUT SYSTEM IS INTENDED FOR BEDSIDE MONITORING OF URINE OUTPUT FOR ANY PATIENT WITH AN INDWELLING, UROLOGICAL CATHETER, DRAINAGE TUBING AND COLLECTION BAG, TYPICALLY IN CRITICAL CARE SETTINGS OR WHERE CLOSE MONITORING OF URINE OUTPUT IS DESIRED. THE BD SENSICA¿ URINE OUTPUT SYSTEM IS ALSO INTENDED FOR MONITORING CORE BLADDER TEMPERATURE WHEN USED WITH THE BD SENSICA¿ TEMPERATURE MONITOR MODULE AND A BD 400 SERIES TEMPERATURE-SENSING FOLEY CATHETER OR EQUIVALENT. CONTRAINDICATIONS: THERE ARE NO KNOWN CONTRAINDICATIONS FOR USE WITH PATIENTS WHO HAVE INDWELLING, UROLOGICAL CATHETERS IN PLACE. WARNINGS: MEDICAL ELECTRICAL EQUIPMENT REQUIRES SPECIAL PRECAUTIONS REGARDING ELECTROMAGNETIC COMPATIBILITY (EMC) AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO THE EMC INFORMATION PROVIDED IN THE CHARTS AT THE END OF THESE INSTRUCTIONS FOR USE. PORTABLE AND MOBILE RADIOFREQUENCY (RF) COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL ELECTRICAL EQUIPMENT. THE BD SENSICA¿ URINE OUTPUT SYSTEM SHOULD NOT BE STACKED WITH OTHER EQUIPMENT. THE BD SENSICA¿ URINE OUTPUT SYSTEM HAS A DEGREE OF PROTECTION AGAINST ELECTRIC SHOCK OF APPLIED PARTS. DO NOT IMMERSE OR SUBMERGE THE BD SENSICA¿ URINE OUTPUT STAND, DISPLAY MONITOR, RING, TEMPERATURE MONITOR MODULE, TEMPERATURE SENSOR CABLES, OR MONO PLUG ADAPTER IN WATER OR OTHER LIQUIDS. DO NOT POUR LIQUIDS OVER THE DISPLAY MONITOR, RING, TEMPERATURE MONITOR, CABLES OR MONO PLUG ADAPTER. IF LIQUIDS ACCIDENTALLY SPILL ONTO THE DEVICE(S), WIPE OFF LIQUID WITH SOFT CLOTH AS SOON AS POSSIBLE. SEE SECTION 14 FOR COMPLETE CARE INSTRUCTIONS. THIS DEVICE IS NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE MIXTURES. THIS DEVICE IS NOT SUITABLE FOR USE IN OXYGEN RICH ENVIRONMENTS. THE USE OF TRANSDUCERS AND CABLES OTHER THAN THOSE SPECIFIED, WITH THE EXCEPTION OF TRANSDUCERS AND CABLES SOLD AS REPLACEMENT PARTS, MAY RESULT IN INCREASED EMISSIONS OR DECREASED IMMUNITY OF THE BD SENSICA¿ URINE OUTPUT SYSTEM. THE BD SENSICA¿ URINE OUTPUT SYSTEM IS DESIGNED TO BE USED WITH ANY BARD¿ FOLEY CATHETER AND UROLOGICAL DISPOSABLES CONNECTED TO ANY STANDARD URINE DRAINAGE BAG (2000 OR 2500 ML), WITH OR WITHOUT A URINE METER. IF USING A 400-SERIES TEMPERATURE-SENSING FOLEY CATHETER, SEE PRODUCT INSTRUCTIONS FOR USE FOR MRI COMPATIBILITY. IF USING A 400-SERIES TEMPERATURE-SENSING FOLEY CATHETER, THE OPERATOR IS RESPONSIBLE FOR ENSURING THE COMPATIBILITY OF THE TEMPERATURE SENSING FOLEY CATHETER TO THE BD SENSICA¿ URINE OUTPUT SYSTEM AND ITS ACCESSORIES. INCOMPATIBLE COMPONENTS CAN RESULT IN DEGRADED PERFORMANCE. NO MODIFICATION OF THE BD SENSICA¿ URINE OUTPUT SYSTEM OR ANY OF ITS ACCESSORIES IS ALLOWED. PRECAUTIONS: DURING SYSTEM START UP AND IN GENERAL PRACTICE, PLUG THE BD SENSICA¿ URINE OUTPUT SYSTEM INTO A WALL POWER SUPPLY WHENEVER POSSIBLE. THE SYSTEM SCREEN WILL DIM WHEN THE SYSTEM IS UNPLUGGED TO MAXIMIZE BATTERY LIFE. AFTER USING THE SYSTEM ON BATTERY BACK-UP, PLUG IT BACK INTO THE WALL POWER SUPPLY FOR RECHARGING AND TO AVOID SYSTEM SHUT DOWN DUE TO A DRAINED BATTERY. AFTER REMOVING A PATIENT FROM URINE OUTPUT MONITORING WITH THE BD SENSICA¿ URINE OUTPUT SYSTEM, SHUT DOWN THE SYSTEM COMPLETELY TO AVOID BATTERY DRAINAGE WHILE NOT IN USE. LOW PRIORITY ALARMS MAY INTERRUPT URINE OUTPUT MONITORING AND REQUIRE USER ACTION TO RESUME NORMAL URINE OUTPUT MONITORING FUNCTIONS." H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT: THE ACTUAL/SUSPECTED DEVICE WAS INSPECTED.
Description of Event or Problem
IT WAS REPORTED THAT THE TOUCHSCREEN CALIBRATION FAILED IN SENSICA DEVICE.
Additional Manufacturer Narrative
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.
Description of Event or Problem
IT WAS REPORTED THAT THE TOUCHSCREEN CALIBRATION FAILED IN SENSICA DEVICE.