MalfunctionIOY

SUPPORT ARM 176

Received Feb 21, 2023 · Event occurred Jan 5, 2023

Report 8010042-2023-00377 · MDR key 16408049

Device

Generic name

Support, Arm

Model number

SUPPORT ARM 176

Catalog number

6405976

Product problems

  • Fracture
  • Fracture

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION OF THE REPORTED ISSUE HAS BEEN FINALIZED. IT WAS REPORTED THAT THE VENTILATOR'S SUPPORT ARM USED FOR HOLDING THE PATIENT CIRCUIT BROKE. PATIENT INVOLVEMENT IS UNKNOWN. THE REPORTED ISSUE WITH BROKEN PART WAS CONFIRMED IN PROBLEM DESCRIPTION AND ADDITIONAL INFORMATION, WHERE IT WAS STATED THAT THE SUPPORT ARM BROKE AT THE CLAMP ATTACHMENT. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGNED ACCORDING TO STANDARD. THE BROKEN SUPPORT ARM WAS NOT RETURNED TO US FOR FURTHER INVESTIGATION. THE ROOT CAUSE OF THE BROKEN SUPPORT ARM HAS NOT BEEN CONCLUSIVELY DETERMINED. HOWEVER, THE BRAKING POINT INDICATES THAT IT MOST LIKELY HAS BEEN EXPOSED TO A MECHANICAL FORCE GREATER THAN IT IS DESIGNED TO SUSTAIN.

Description of Event or Problem

IT WAS REPORTED THAT THE VENTILATOR'S SUPPORT ARM USED FOR HOLDING THE PATIENT CIRCUIT BROKE. PATIENT INVOLVEMENT IS UNKNOWN. MANUFACTURER'S REF #: (B)(4).