Received Feb 17, 2023 · Event occurred Jan 30, 2023

Report 1823260-2023-00500 · MDR key 16398333

Device

Generic name

Phosphomolybdate (colorimetric), Inorganic Phosphorus

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

03183793122

Lot number

67081901

Product problems

  • Low Test Results
  • Low Test Results

Patient

60 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH PHOS2 PHOSPHATE (INORGANIC) VER.2 ON A COBAS 6000 C (501) MODULE (SERIAL NUMBER (B)(4) ). NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN A PHOSPHATE VALUE OF 1.0 MG/DL WITH A DATA FLAG. THE SAMPLE WAS REPEATED, RESULTING IN A VALUE OF 3.2 MG/DL AND THIS VALUE WAS DEEMED CORRECT AND REPORTED OUTSIDE OF THE LABORATORY.

Additional Manufacturer Narrative

THE CALIBRATION PERFORMED ON 29-JAN-2023 WAS OK AND THERE WERE NO ALARMS. THE SAME REAGENT PACK WAS CALIBRATED ON 30-JAN-2023 AND THERE WERE VARIATIONS IN THE CALIBRATION SIGNALS. ONE LEVEL OF QUALITY CONTROL SHIFTED OUTSIDE OF RANGE ON 29-JAN-2023 AND A SECOND LEVEL OF CONTROL SHIFTED OUTSIDE OF RANGE ON 30-JAN-2023. THE FIELD SERVICE ENGINEER CHECKED THE SYSTEM AND COULD NOT FIND ANYTHING WRONG. THE CUSTOMER REPLACED THE REAGENT AND NO ISSUES WERE OBSERVED. THE CUSTOMER BELIEVES THE REAGENT "STAYED OUT FOR A WHILE". THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.