inforMED
MalfunctionIOY

SUPPORT ARM 176

Received Feb 14, 2023 · Event occurred Jan 20, 2023

Report 8010042-2023-00331 · MDR key 16373265

Device

Generic name

Support, Arm

Model number

SUPPORT ARM 176

Catalog number

6405976

Product problems

  • Fracture
  • Fracture

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE SUPPORT ARM HOLDING THE PATIENT TUBES BROKE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF #: (B)(4).

Additional Manufacturer Narrative

THE SUPPORT ARM WAS REPAIRED BY OUR FIELD SERVICE ENGINEER AND WAS CLEARED FOR CLINICAL USE. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGNED ACCORDING TO STANDARD. THE ROOT CAUSE OF THE BROKEN SUPPORT ARM HAS NOT BEEN CONCLUSIVELY DETERMINED. HOWEVER, THE BREAKING POINT INDICATES THAT IT MOST LIKELY HAS BEEN EXPOSED TO A MECHANICAL FORCE GREATER THAN IT IS DESIGNED TO SUSTAIN.

Description of Event or Problem

MANUFACTURER'S REF #: (B)(4).