inforMED
MalfunctionNLH

NA

Received Feb 13, 2023 · Event occurred Nov 28, 2022

Report 0002090040-2023-00001 · MDR key 16362831

Device

Generic name

Catheter, Recording, Electrode, Reprocessed

Model number

200574

Catalog number

200574RR

Lot number

4848945

Product problems

  • Noise, Audible
  • Noise, Audible

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE DEVICE HAD A BUZZING SOUND. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS FIVE MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO STRYKER SUSTAINABILITY SOLUTIONS FOR EVALUATION. UPON INSPECTION OF THE RECEIVED COMPLAINT DEVICE, THE DEVICE DID NOT MEET THE CURVE SPECIFICATION. A REVIEW OF THE DHR SUPPORTS THAT THE DEVICE MET ALL INSPECTION AND TEST CRITERIA PRIOR TO RELEASE FROM STRYKER. THE MOST LIKELY ROOT CAUSE IS MISHANDLING SUBSEQUENT TO DISTRIBUTION, INCLUDING SHIPPING/STORAGE CONDITIONS OR IMPROPER MANIPULATION OF THE CATHETER. THE INSTRUCTIONS FOR USE (IFU) STATE: DO NOT EXERT EXCESSIVE PRESSURE DURING PLACEMENT OF CATHETER IF UNKNOWN RESISTANCE IS ENCOUNTERED. DO NOT ATTEMPT TO USE THE REPROCESSED EP CATHETER PRIOR TO COMPLETELY READING AND UNDERSTANDING THE DIRECTIONS FOR USE. AVOID MANUAL PRE-BENDING OF DISTAL CURVE, AS THIS MAY DAMAGE STEERING MECHANISM OF STEERABLE CATHETER THE REPORTED EVENT WILL CONTINUE TO BE MONITORED THROUGH POST-MARKET SURVEILLANCE.