inforMED
MalfunctionMLM

DIMENSION®

Received Feb 8, 2023 · Event occurred May 26, 2022

Report 2517506-2023-00029 · MDR key 16330244

Device

Generic name

Dimension® (tac) Tacrolimus Flex® Reagent Cartridge

Model number

10700795

Catalog number

DF207 SMN 10700795

Lot number

GA2286

Product problems

  • Imprecision
  • Imprecision

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER INFORMED SIEMENS OF IMPRECISE ELEVATED QUALITY CONTROL (QC) RESULTS FOR TACROLIMUS (TAC) WITH LOT GA2286 ON A DIMENSION EXL 200 SYSTEM. NO DATA WAS PROVIDED. NO PATIENT SAMPLES WERE IMPACTED BY THIS ISSUE. THERE IS NO INDICATION OF DELAY OF PATIENT RESULT REPORTING. THERE IS NO INDICATION THAT PATIENT TREATMENT WAS ALTERED, PRESCRIBED OR DELAYED ON THE BASIS OF TAC QUALITY CONTROL RESULTS OUTSIDE OF CUSTOMER ACCEPTANCE RANGE.

Additional Manufacturer Narrative

AN OUTSIDE OF UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REMOTE SERVICES CENTER (RSC) REGARDING IMPRECISION FOR QUALITY CONTROL (QC) SAMPLES WITH DIMENSION TACROLIMUS (TAC) LOT GA2286 ON A DIMENSION EXL 200 SYSTEM. NO PATIENT SAMPLES WERE IMPACTED BY THIS ISSUE. SIEMENS HEALTHCARE DIAGNOSTICS CONCLUDED THE INVESTIGATION OF THE EVENT. SIEMENS CONFIRMED IMPRECISION FOR QUALITY CONTROL (QC) AND PATIENT SAMPLES WITH DIMENSION TACROLIMUS (TAC) LOTS GA2286, GA3047 AND GA3171. URGENT MEDICAL DEVICE CORRECTION DC-23-03.A.US DATED 30-JAN-2023 WAS SENT TO ALL UNITED STATES (US) CUSTOMERS WHO HAD BEEN SHIPPED THE IMPACTED LOTS. URGENT FIELD SAFETY NOTICE (UFSN) DC-23-03.A.OUS DATED JANUARY 2023 WAS SENT TO OUTSIDE THE US (OUS) CUSTOMERS WHO HAD BEEN SHIPPED THE IMPACTED LOTS. CUSTOMERS WERE ADVISED TO DISCONTINUE USE AND DISCARD THE DIMENSION TACROLIMUS LOTS GA2286, GA3047 AND GA3171. ROOT CAUSE INVESTIGATION IS IN PROGRESS AND CORRECTIVE ACTION UPDATES WILL BE PROVIDED IN MONTHLY STATUS UPDATE REPORTS TO THE FDA AS PER 21 CFR PART 806.10.

Remedial action

  • Recall