inforMED
MalfunctionNFB

CANNULA SOFT ADULT W/U-CON 7FT TB 50/CS

Received Feb 6, 2023 · Event occurred Jan 12, 2023

Report 8030673-2023-00296 · MDR key 16305622

Device

Generic name

Conserver, Oxygen

Catalog number

SFT2699

Lot number

0004232076

Product problems

  • Material Deformation
  • Material Deformation

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEVICE EVALUATION: G3, G6, H2, H6 AND H10. VYAIRE DETERMINED THAT MANUFACTURING PERSONNEL MAY BE RELATED WITH THE REPORTED DEFECT SINCE IT WAS REGISTERED THAT NEW PERSONNEL WAS WORKING ON THE LINE AT THE TIME THE FG PART NUMBER SFT2699 WAS BEING MANUFACTURED. THIS BRING US THAT NEW PERSONNEL MAY NOT HAVE FOLLOWED CORRECTLY THE PROCEDURE ACCORDING TO SPM 001309 ETAL ON HOW TO PROPERLY ASSEMBLY THE FG SFT2699, HOWEVER IS AND ISOLATED CASE SINCE WE HAVE NOT RECEIVED ANY COMPLAINTS ABOUT THIS ISSUE IN THE PAST. IN ADDITION, PFMEA-14A-005 WAS REVIEWED AND WE HAVE CONTROLS FOR THIS TYPE OF ISSUES. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative

AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. NO ROOT CAUSE HAS BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE."

Description of Event or Problem

IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SFT2699 CANNULA SOFT ADULT W/U-CON 7FT TB 50/CS CANNULA IS LOOPED SO THE PATIENT JUST BREATHES IN THEIR OWN AIR. THERE IS NO INFORMATION ON THE PATIENT OUTCOME.