inforMED
MalfunctionJHJ

ICON 20 HCG PREGNANCY TEST

Received Feb 1, 2023 · Event occurred Dec 15, 2022

Report MW5114626 · MDR key 16282809

Device

Generic name

Visual, Pregnancy Hcg, Prescription Use

Model number

395097

Catalog number

B3167-50

Lot number

032H11

Product problems

  • False Positive Result

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

HIGH RATE OF FALSE POSITIVE RESULTS ON BECKMAN COULTER ICON 20 HCG KITS, LOT 032H11. FALSE POSITIVES HAVE OCCURRED ON EIGHT DIFFERENT PATIENTS. DATES OF OCCURRENCE WERE (B)(6) 2022, (B)(6) 2022, (B)(6) 2022, (B)(6) 2022, (B)(6) 2022, (B)(6) 2022, (B)(6) 2023, AND (B)(6) 2023. RESULTS WERE DETERMINED TO BE FALSE POSITIVES BASED ON REPEATING THE URINE ON ANOTHER LOT OF ICON 20 HCG KITS, RUNNING THE PATIENTS' SERUM HCG (BECKMAN ACCESS 2), AND/OR ULTRASOUND. PATIENT AGES RANGE FROM 17 TO 45 YEARS. PATIENTS INCLUDE 7 CIS FEMALES AND 1 TRANSGENDER MALE. APPROXIMATELY 451 TESTS HAVE BEEN PERFORMED ON THIS LOT BETWEEN TWO LOCATIONS, (B)(6). APPROXIMATE FALSE POSITIVE RATE IS (B)(4). REFERENCE REPORTS: MW5114623, MW5114624, MW5114625, MW5114627, MW5114628, MW5114629, MW5114630.