Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE BLADE EDGES HAD PERFORATED THE STERILE PACKAGING. NO ADVERSE CONSEQUENCES WERE REPORTED.
Additional Manufacturer Narrative
THERE WERE A NUMBER OF ITEMS ASSOCIATED WITH THIS COMPLAINT. IT CAN BE CONFIRMED FROM VISUAL INSPECTION THAT PRODUCT PACKAGING IS DAMAGED. OTHER COMPLAINTS HAVE BEEN RECEIVED REPORTING SIMILAR EVENTS. A SECONDARY POLYBAG HAS BEEN INTRODUCED INTO THE PACKAGING CONFIGURATION OF THESE PRODUCTS TO ADDRESS THIS ISSUE.