BD BBL¿ TAXO¿ DIFFERENTIATION DISCS FOR V FACTOR

Received Jan 19, 2023 · Event occurred Jan 10, 2023

Report 1119779-2023-00045 · MDR key 16209274

Device

Generic name

Discs, Strips And Reagents, Microorganism Differentiation

Model number

231727

Catalog number

231727

Lot number

2056543

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT 60 BD BBL¿ TAXO¿ DIFFERENTIATION DISCS FOR V FACTOR DURING VALIDATION NO GROWTH OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NO GROWTH OF HAEMOPHILUS INFLUENZAE AT THE EDGE OF 60 DISC WITH THE USE OF A COS + FACTOR V MEDIUM DURING VALIDATION NO GROWTH , ARE TESTING NOW WITH OTHER BATCH , TOMORROW NOON 11/1/2023 12H I HAVE TO CALL THE CUSTOMER FOR RESULTS OTHER BATCH.

Additional Manufacturer Narrative

H6 INVESTIGATION SUMMARY: A COMPLAINT INVESTIGATION DUE TO NO GROWTH FOR TAXO V FACTORS CATALOG 231727 BATCH NO.: 6924935 WAS PERFORMED ON RETENTION SAMPLES. RETURNED GOODS WERE NOT RECEIVED FROM CUSTOMER. THE INVESTIGATION REQUIRED TO TEST PRODUCT FOR PERFORMANCE, VISUAL INSPECTION, AND BATCH RECORD REVIEW. RETENTION SAMPLES PERFORMED AS EXPECTED. NO DISCREPANCIES OBSERVED DURING THE VISUAL INSPECTION. BATCH RECORD REVIEW DID NOT SHOW ANY EVIDENCE OF CUSTOMER CLAIM. NO CORRECTIVE ACTION IS REQUIRED BASED ON INFORMATION EVALUATED AND SATISFACTORY RESULTS OBTAINED DURING THE QC TEST. COMPLAINT IS NOT CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED.

Additional Manufacturer Narrative

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED THAT 60 BD BBL¿ TAXO¿ DIFFERENTIATION DISCS FOR V FACTOR DURING VALIDATION NO GROWTH OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NO GROWTH OF HAEMOPHILUS INFLUENZAE AT THE EDGE OF 60 DISC WITH THE USE OF A COS + FACTOR V MEDIUM DURING VALIDATION NO GROWTH , ARE TESTING NOW WITH OTHER BATCH , TOMORROW NOON 11/1/23 12H I HAVE TO CALL THE CUSTOMER FOR RESULTS OTHER BATCH.