ELECSYS PROBNP II
Received Jan 17, 2023 · Event occurred Dec 30, 2022
Report 1823260-2023-00145 · MDR key 16185771
Device
Generic name
Test,natriuretic Peptide
Manufacturer
Roche DiagnosticsModel number
NACatalog number
09315268190
Lot number
63046801
Product problems
- High Readings
- Non Reproducible Results
- High Readings
- Non Reproducible Results
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE LAST CALIBRATION PERFORMED ON (B)(6) 2022 WAS WITHIN EXPECTATIONS. UPON REVIEW OF QUALITY CONTROL DATA, CONTROLS WERE LAST TESTED ON (B)(6) 2022 AND NOT ON THE DATE OF THE EVENT. PER PRODUCT LABELING: "CONTROLS FOR THE VARIOUS CONCENTRATION RANGES SHOULD BE RUN INDIVIDUALLY AT LEAST ONCE EVERY 24 HOURS WHEN THE TEST IS IN USE, ONCE PER REAGENT KIT, AND FOLLOWING EACH CALIBRATION." BASED ON THE INFORMATION, A GENERAL REAGENT ISSUE CAN MOST LIKELY BE EXCLUDED. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED A QUESTIONABLE RESULT FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS PROBNP II IMMUNOASSAY VER. 2.1 ON A COBAS 6000 E 601 MODULE (SERIAL NUMBER (B)(4)). NO INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. NO UNITS OF MEASURE WERE PROVIDED. THE SAMPLE INITIALLY RESULTED IN A PROBNP VALUE OF > 35000. THE SAMPLE WAS REPEATED USING THE BIOMEREUX VIDAS METHOD, RESULTING IN A VALUE OF 15000. THE CUSTOMER THEN REPLACED THE PROBNP REAGENT PACK ON THE E601 ANALYZER AND REPEATED THE SAMPLE ON THE E601, RESULTING IN A VALUE THAT MATCHED THE 15000 VALUE MEASURED WITH THE VIDAS METHOD.