inforMED
MalfunctionKPZ

VERIFLEX

Received Apr 2, 1998 · Event occurred Oct 15, 1997

Report 1121753-1998-00013 · MDR key 161701

Device

Generic name

Record And Verify System

Manufacturer

Nucletron Bv

Model number

VERSION 2.04

Catalog number

109.020

Lot number

NA

Product problems

  • Computer Software Problem
  • Application Program Problem: Parameter Calculation Error

Patient

NA

  • Overdose

Narrative

Description of Event or Problem

DURING THE FIRST TREATMENT OF A PT, TWO OF TWO FIELDS WERE TREATED. THE THERAPIST THEN PREPARED TO TAKE A PORT FILM. THE ACCELERATOR ATTEMPTED TO DELIVER 250 MU, BUT WHEN THE THERAPIST REALIZED THE ERROR, 100 MU HAD BEEN DELIVERED. THIS EXCEEDED THE TOTAL PRESCRIBED MU BY 100. IF ESC KEY IS PRESSED ON THE VERIFLEX KEYBOARD, AFTER RAD-ON AND BEFORE RAD-OFF, VERIFLEX WILL FAIL TO RECORD TREATMENT AND WILL NOT REGISTER THE TREATMENT AS BEING COMPLETE. IT IS THEN POSSIBLE TO RE-TREAT THE SAME FIELD.

Remedial action

  • Notification