inforMED
MalfunctionLMB

GUARDIAN

Received Feb 8, 2010 · Event occurred Jul 8, 2008

Report 1064427-2009-00030 · MDR key 1610478

Device

Model number

NA

Catalog number

250001009

Lot number

NA

Product problems

  • Computer Operating System Problem

Patient

UNK

  • No Information

Narrative

Additional Manufacturer Narrative

THERE WAS NO PATIENT INVOLVED, THUS NO PATIENT DATA EXISTS. THE INITIAL REPORTER WAS AN IT ENGINEER. THERE IS NO EXPIRATION DATE FOR THIS PRODUCT. THIS IS A REMOTE DATA BACKUP SERVICE OF WHICH THERE IS NO OPERATOR. ACTUAL DEVICE WAS EVALUATED REMOTELY. THERE IS NO KNOWN INFORMATION ON THE MANUFACTURE DATE AT THIS TIME. EVALUATION SUMMARY - THE SYSTEM WAS EVALUATED REMOTELY ON 08/19/2008 AND FOUND TO HAVE TOO MANY FILES FOR GUARDIAN TO SUPPORT. WHEN THERE ARE TOO MANY FILES PRESENT, GUARDIAN CANNOT PERFORM BACKUPS DURING THE HOURS ALLOTTED TO THIS TASK SO, IT WILL CONTINUE TO RUN THROUGH BUSINESS HOURS THUS SLOWING DOWN THE SITE'S SYSTEM. AT THIS POINT GUARDIAN IS SHUT DOWN WHICH CREATES A POTENTIAL FOR PATIENT DATA LOSS. HOWEVER, THERE IS NO EVIDENCE OF ANY PATIENT DATA LOSS IN THIS PARTICULAR INSTANCE. THIS IS NOT A SINGLE USE DEVICE. (B)(4)

Description of Event or Problem

THE GUARDIAN REPLIWEB IS UNABLE TO HANDLE THE SITE FILES.

Remedial action

  • Recall