INSULIN PUMP CONTROL DEVICE (PDM)

Received Dec 22, 2022 · Event occurred Nov 12, 2022

Report MW5113995 · MDR key 16043560

Device

Generic name

Continuous Glucose Monitor Informed Insulin Dose Calculator

Product problems

  • Inadequate User Interface
  • Application Program Problem: Medication Error

Patient

50 YR · Male

  • Hypoglycemia
  • Hypoglycemia

Narrative

Description of Event or Problem

INSULIN DISPENSED UNDER THE AUTOMATION MODE OF INSULET'S OMNIPOD 5 INSULIN PUMP IS NOT ACCOUNTED IN THE INSULIN ON BOARD TALLY THAT IS PROVIDED BY THE INSULIN PUMP CONTROL DEVICE (PDM) WHICH IS ALSO AN INSULET MANUFACTURED MEDICAL DEVICE AND IS A SEPARATE DEVICE FROM THE OMNIPOD INSULIN PUMP POD. THE INSULIN ON BOARD TALLY IS RELIED UPON BY THE INSULIN PUMP OPERATOR TO DETERMINE APPROPRIATE INSULIN BOLUS DOSAGE WHEN A BOLUS IS NEEDED FOR MEALS OR HIGH GLUCOSE LEVEL CORRECTION. IN MY EPISODE, THE PUMP HAD DISPENSED NEARLY 3 UNITS OF INSULIN OVER THE COURSE OF AN HOUR ACROSS 5 MINUTE INCREMENTS. THIS WAS NOT REFLECTED IN THE INSULIN ON BOARD TALLY. SO UPON BEING WOKEN BY A HIGH GLUCOSE ALARM, I ORDERED THE PDM TO BOLUS A DOSAGE OF INSULIN RELYING ON THE INSULIN ON BOARD TALLY VERSUS THE CURRENT GLUCOSE STATUS. THE BOLUS PLUS THE EXISTING INSULIN FROM THE PUMPS AUTOMATION IN MY SYSTEM WAS TOO MUCH AND I WAS AWOKEN LATER IN THE EVENING WITH A LOW GLUCOSE ALARM FINDING THAT MY GLUCOSE WAS QUICKLY DROPPING BELOW 47.