inforMED
OtherGDY

LAP SPONGES

Received Jan 28, 2010 · Event occurred Aug 3, 2009

Report 2083545-2010-00001 · MDR key 1603374

Device

Generic name

Lap Sponges

Catalog number

340463

Lot number

7408

Product problems

  • Device Contamination with Chemical or Other Material

Patient

Not reported

  • No Information

Narrative

Additional Manufacturer Narrative

NO RETURNED SAMPLE FROM THE (B)(6) HOSPITAL OF (B)(6) WAS AVAILABLE IN CONNECTION TO THE MEDWATCH REPORT. SOME STERILE LAP SPONGE PACK SAMPLES FROM THE SAME LOT NUMBER WERE RETURNED BY THE (B)(6) HOSPITAL RECENTLY. HOWEVER, WE CANNOT BE SURE THAT THE SAMPLES WERE RELATED TO THE RECENT MEDWATCH REPORT. SAMPLES FROM LOT NUMBER 7408 PREVIOUSLY RETURNED FROM THE (B)(6) HOSPITAL WERE CAREFULLY EXAMINED. NO LINT OR LOOSE ATTACHED THREAD THAT MIGHT BE DESCRIBED AS "LINT" WAS FOUND IN THE RETURNED UNOPENED PACKAGES OF THE STERILE LAP SPONGES. SOME PACKAGES WERE ALSO OPENED TO CHECK THE LAP SPONGES, NO NONCONFORMING CONDITION WAS DISCOVERED. WE ALSO CHECKED THE PRODUCTION AND QUALITY RECORDS (INCLUDING THE USP GAUZE MONOGRAPH TEST RESULTS) FOR THE LOT NUMBER IN QUESTION, AND NO NONCONFORMING CONDITION WAS FOUND. WE ALSO CHECKED THE CUSTOMER COMPLAINT TRENDING RECORDS, AND NO OTHER SIMILAR COMPLAINT WAS REPORTED BY OTHER CUSTOMERS ON THIS LOT NUMBER OF LAP SPONGES. (B)(4). HOWEVER, THIS DOES NOT AFFECT THE SAFETY AND INTENDED USE OF THE PRODUCTS.

Description of Event or Problem

EVENT DESC: UNOPENED PACKAGES OF STERILE LAP SPONGES CONTAIN LINT. DEVICE USAGE PROBLEM: OTHER.