inforMED
MalfunctionKZF

RITTER EXAM TABLE

Received Dec 19, 2022 · Event occurred Dec 1, 2022

Report MW5113861 · MDR key 16016074

Device

Generic name

Device, Medical Examination, Ac Powered

Model number

104

Product problems

  • Temperature Problem

Patient

35 YR · Female

  • Superficial (First Degree) Burn
  • Superficial (First Degree) Burn

Narrative

Description of Event or Problem

PATIENT RECEIVED A BURN WHEN A VAGINAL SPECULUM WAS REMOVED FROM HEALING DRAWER IN EXAM TABLE. PATIENT QUICKLY ALERTED PROVIDER OF WARMTH OF INSTRUMENT AND INSTRUMENT WAS REMOVED. TABLE WAS TURNED OFF AND EQUIPMENT REMOVED. THE EXAM TABLE IS A RITTER MODEL 104 PURCHASED IN FEB 2001, ASSET # (B)(4). TABLES ARE INSPECTED ANNUALLY, NEXT DUE FOR INSPECTION FEB 2023. THE TEMPERATURE OF DRAWER WAS REPORTED TO BE 200 DEGREES F.