inforMED
MalfunctionCBF

SERVO-U

Received Dec 19, 2022 · Event occurred Dec 1, 2022

Report 8010042-2022-02259 · MDR key 16004148

Device

Generic name

Chamber, Hyperbaric

Model number

SERVO-U

Catalog number

6694800

Product problems

  • Contamination /Decontamination Problem
  • Contamination /Decontamination Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE VENTILATOR WAS INVESTIGATED ON-SITE AND FOUND PRESENCE OF LIQUID/MEDICINES SPILLED INTO THE PRESSURE TRANSDUCER PRINTED CIRCUIT BOARDS (PCBS). THE PRESSURE TRANSDUCER PCBS WER DEFECTIVE AND REPLACED. AFTER REPLACEMENT, THE VENTILATOR PASSED ALL FUNCTIONAL AND SAFETY TESTS PER FACTORY SPECIFICATIONS, RETURNED TO CUSTOMER AND CLEARED FOR CLINICAL USE. THE PROVIDED PICTURES CONFIRMED THE REPORTED LIQUID/MEDICINES INGRESS INTO THE PCBS. THE PROVIDED PICTURES CONFIRMED THE LIQUID ON THE PCBS. INGRESS OF LIQUID/CORROSIVE SUBSTANCE ON PC BOARDS CAN CAUSE FAILURE/SHORT CIRCUITS, WHICH MIGHT LEAD TO A VENTILATOR MALFUNCTION. THE PROVIDED PICTURES RAVELED SIGNS OF LIQUID SPILL. THERE IS NO INFORMATION ON HOW THE LIQUID SPILL ENTERED THE VENTILATOR OR WHAT KIND OF LIQUID SPILL IT WAS. OUR CONCLUSION IS THAT THE REPORTED PROBLEM WAS CAUSED BY DEFECTIVE PRESSURE TRANSDUCER PCBS AS A RESULT OF WATER INGRESS INTO THE VENTILATOR.

Description of Event or Problem

IT WAS REPORTED THAT THE VENTILATOR FAILED PRESSURE TRANSDUCER TEST DURING PRE-USE CHECK. THE VENTILATOR WAS CONTAMINATED WITH WATER/LIQUID. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF.#: (B)(4).