inforMED
MalfunctionMSW

IMMULITE 2000 XPI THYROGLOBULIN

Received Dec 9, 2022 · Event occurred Oct 18, 2022

Report 2432235-2022-00330 · MDR key 15950418

Device

Generic name

Immulite 2000 Thyroglobulin

Model number

N/A

Catalog number

10381648

Lot number

433

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN OUTSIDE OF U.S. (OUS) CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER TO REPORT IMPRECISION ON QUALITY CONTROL SAMPLES ON THE IMMULITE 2000 SYSTEMS WHEN USING THYROGLOBULIN KIT LOT: 435. SIEMENS HEALTHCARE DIAGNOSTICS FURTHER INVESTIGATED THE ISSUE AND HAS CONFIRMED THE POTENTIAL FOR FUNCTIONAL SENSITIVITY TO NOT MEET INSTRUCTIONS FOR USE CLAIMS WITH THYROGLOBULIN KIT LOTS: D431-D436 (US) AND KIT LOTS: 431-436 (OUTSIDE THE US). INVESTIGATION HAS ALSO DEMONSTRATED THAT IMMULITE THYROGLOBULIN CONTROL MODULE CONTROL LEVEL ONE MAY RESULT OUTSIDE OF PUBLISHED RANGES. PER GOOD LABORATORY PRACTICE, PATIENT RESULTS ARE NOT REPORTED WHEN CONTROLS RESULT OUT OF RANGE. WHEN CONTROL RESULTS ARE IN RANGE, USERS MAY OBSERVE INCREASED IMPRECISION WITH LOW LEVEL PATIENT SAMPLES. AN URGENT MEDICAL DEVICE CORRECTION (UMDC, LETTER IMC23-04.A.US) WAS SENT TO U.S. CUSTOMERS (B)(6) 2022 AND AN URGENT FIELD SAFETY NOTICE (UFSN, LETTER IMC23-04.A.OUS) WAS SENT TO OUTSIDE THE US (OUS) CUSTOMERS ON (B)(6) 2022. THE UMDC AND UFSN EXPLAIN THE POTENTIAL FOR FUNCTIONAL SENSITIVITY NOT MEETING IFU CLAIMS AND INCREASED IMPRECISION WITH PATIENT SAMPLES WITH LOW LEVEL PATIENT SAMPLES.

Description of Event or Problem

THE CUSTOMER INDICATED IMPRECISION ON QUALITY CONTROL SAMPLES ON THE IMMULITE 2000 SYSTEMS WHEN USING THYROGLOBULIN KIT LOT 435. THERE WERE NO DISCORDANT PATIENT RESULTS PROVIDED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE IMPRECISION ON QUALITY CONTROL SAMPLES.

Remedial action

  • Recall