inforMED
InjuryNXO

FECAL EXTRACTION DEVICE

Received Dec 9, 2022 · Event occurred Sep 29, 2020

Report 2026994-2022-00001 · MDR key 15950317

Device

Generic name

Calprotectin, Fecal

Model number

504837

Catalog number

504837

Product problems

  • Use of Device Problem
  • Device Handling Problem
  • Use of Device Problem
  • Device Handling Problem

Patient

Unknown

  • Limb Pain
  • Limb Pain

Narrative

Description of Event or Problem

ON (B)(6)2022, THE PATIENT (SPECIAL CHEMISTRY TECHNOLOGIST, HEALTHCARE PROFESSIONAL) FILED A SUPERVISOR'S INCIDENT INVESTIGATION AT LABCORP REPORTING RIGHT ARM HAND PAIN AND NUMBNESS. THE INJURY WAS SUSPECTED TO BE DUE TO THE CONTINUAL REPETITIVE MOTION OF PERFORMING BIOLOGICAL SAMPLE EXTRACTIONS. THE PATIENT ONLY RAN CALPROTECTIN TESTING AND HAD DIFFICULTY USING THE FECAL EXTRACTION DEVICE. THE EMPLOYEE REPORTED HANDLING BETWEEN 600-900 SAMPLES IN A SINGLE SHIFT, USING THE SAME REPETITIVE MANUAL MOTION. IT WAS NOTED THAT THE PATIENT WAS SCHEDULED FOR FURTHER MEDICAL EVALUATION BY ORTHOPEDIC STAFF AT THAT TIME. THIS INCIDENT WAS NOT REPORTED TO INOVA DIAGNOSTICS AT THIS TIME. ON (B)(6)2022, INOVA DIAGNOSTICS WAS CONTACTED BY THE LABCORP OF AMERICA CUSTOMER AGAIN TO REQUEST SPECIAL TOOLS TO AID IN OPENING A HIGH VOLUME OF FECAL EXTRACTION DEVICES. THEY REPORTED THAT THE FECAL EXTRACTION DEVICE VOLUME HAD GROWN AS THE LABORATORY IS PROCESSING OVER 25,000 SAMPLES PER MONTH. THERE WERE NO PERMANENT WRIST INJURIES OR CARPAL TUNNEL DIAGNOSES REPORTED WITH CURRENT USERS, HOWEVER, THE INJURY THAT OCCURRED IN 2020 WAS SHARED WITH INOVA DIAGNOSTICS AT THIS TIME AS EVIDENCE OF A PAST INJURY WHICH REQUIRED INTERVENTION (SEE (B)(6) 2020 DESCRIPTION, ABOVE). THE CUSTOMER SPECIFIED THAT THE INJURY WAS LATER DIAGNOSED AS CARPAL TUNNEL SYNDROME AND THE TECHNOLOGIST WAS TRANSFERRED TO ANOTHER DEPARTMENT TO PREVENT FURTHER INJURY.

Additional Manufacturer Narrative

THE CUSTOMER WAS REMINDED OF PROPER DEVICE OPENING TECHNIQUE, AS DETAILED IN THE PRODUCT DIRECTION INSERT. THE CUSTOMER WAS REMINDED TO FOLLOW THE INSTRUCTIONS CAREFULLY TO ENSURE PROPER AND COMFORTABLE REMOVAL OF THE DEVICE CAPS. THE CUSTOMER WAS ALSO SENT SPECIAL TOOLS TO ASSIST WITH OPENING THE DEVICE. THE TOOLS ADD AN ADDITIONAL GRIPPING AREA, WHICH REDUCES THE STRAIN ON THE OPERATOR WHEN OPENING A LARGE NUMBER OF DEVICES.

Remedial action

  • Other