TRACTION BOOT ASSEMBLY TO ORTHOPEDIC SURGICAL TABLE
Received Dec 9, 2022 · Event occurred Nov 11, 2022
Report 1043572-2022-00086 · MDR key 15949887
Device
Generic name
Traction Boot Assembly To Orthopedic Surgical Table
Manufacturer
Steris Corporation - MontgomeryModel number
BF538Catalog number
BF538
Product problems
- Unintended Movement
- Unintended Movement
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED THAT DURING A PROCEDURE A PATIENT'S FOOT FELL OUT OF THE TRACTION BOOT ASSEMBLY TO THEIR ORTHOPEDIC SURGICAL TABLE. THE PATIENT'S FOOT NEEDED TO BE REALIGNED IN THE BOOT ASSEMBLY CAUSING A PROCEDURE DELAY. NO REPORT OF INJURY AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Additional Manufacturer Narrative
A STERIS SURGICAL PROJECT MANAGER ARRIVED ONSITE TO INSPECT THE TRACTION BOOT ASSEMBLY TO THEIR ORTHOPEDIC SURGICAL TABLE AND WAS UNABLE TO DUPLICATE THE REPORTED EVENT. THE SURGICAL TABLE AND TRACTION BOOT ASSEMBLY WERE FOUND TO BE OPERATING ACCORDING TO SPECIFICATION AND RETURNED TO SERVICE. NO REPAIRS WERE REQUIRED. THE OPERATOR MANUAL (150832-354) STATES IN SEC. 3.4 "USING CARBON FIBER TRACTION UNIT AND ACCESSORIES; STEP 3. A. APPLY TRACTION BOOT ASSEMBLY (SEE FIGURE 3-4) TO PATIENT'S ANATOMY AND SECURE WITH HOOK AND LOOP FASTENER SYSTEM ATTACHED TO FOOT SUPPORT." A STERIS ACCOUNT MANAGER OFFERED IN-SERVICE TRAINING ON THE PROPER USE AND OPERATION OF THEIR TRACTION BOOT ASSEMBLY. THE USER FACILITY ACCEPTED, HOWEVER, THE DATE OF IN SERVICE TRAINING HAS YET TO BE SCHEDULED. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.