IMMULITE 2000 THYROGLOBULIN
Received Dec 8, 2022 · Event occurred Mar 29, 2022
Report 2432235-2022-00382 · MDR key 15947139
Device
Generic name
Immulite 2000 Thyroglobulin
Manufacturer
Siemens Healthcare Diagnostics Products, LimitedModel number
N/ACatalog number
10381648
Lot number
432
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER REPORTED AN IMPRECISION ON THYROGLOBULIN QUALITY CONTROL (LOT 432) ON AN IMMULITE 2000 INSTRUMENT. DUE TO THE QUALITY CONTROL (QC) ISSUE THE CUSTOMER HAS STOPPED PROCESSING THYROGLOBULIN PATIENT RESULTS. THERE WERE NO PATIENT RESULTS GENERATED OR REPORTED AT THE TIME. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Additional Manufacturer Narrative
AN OUTSIDE OF THE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER (CCC) TO REPORT FAILED QUALITY CONTROL SAMPLES ON THE IMMULITE 2000 WHEN USING THYROGLOBULIN KIT LOT 431. THERE WAS NO PATIENT DATA PROVIDED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FAILED QUALITY CONTROL SAMPLES. SIEMENS HEALTHCARE DIAGNOSTICS FURTHER INVESTIGATED THE ISSUE AND HAS CONFIRMED THE POTENTIAL FOR FUNCTIONAL SENSITIVITY TO NOT MEET INSTRUCTIONS FOR USE CLAIMS WITH THYROGLOBULIN KIT LOTS D431-D436 (US) AND KIT LOTS 431-436 (OUTSIDE THE US). INVESTIGATION HAS ALSO DEMONSTRATED THAT IMMULITE THYROGLOBULIN CONTROL MODULE CONTROL LEVEL ONE MAY RESULT OUTSIDE OF PUBLISHED RANGES. PER GOOD LABORATORY PRACTICE, PATIENT RESULTS ARE NOT REPORTED WHEN CONTROLS RESULT OUT OF RANGE. WHEN CONTROL RESULTS ARE IN RANGE, USERS MAY OBSERVE INCREASED IMPRECISION WITH LOW LEVEL PATIENT SAMPLES. AN URGENT MEDICAL DEVICE CORRECTION (UMDC, LETTER IMC23-04.A.US) WAS SENT TO US CUSTOMERS 12/8/2022, AND AN URGENT FIELD SAFETY NOTICE (UFSN, LETTER IMC23-04.A.OUS) WAS SENT TO OUTSIDE THE US (OUS) CUSTOMERS ON 12/7/2022. THE UMDC AND UFSN EXPLAIN THE POTENTIAL FOR FUNCTIONAL SENSITIVITY NOT MEETING IFU CLAIMS AND INCREASED IMPRECISION WITH PATIENT SAMPLES WITH LOW LEVEL PATIENT SAMPLES.
Remedial action
- Recall