LOCK/NEUTRAL GUIDE F/4.5MM VA LCP CRVD CONDYLAR AIMING ARM
Received Dec 2, 2022 · Event occurred Nov 10, 2022
Report 8030965-2022-10597 · MDR key 15907385
Device
Generic name
Condylar Plate Fixation Implant
Manufacturer
Synthes GmbhModel number
03.231.007Catalog number
03.231.007
Lot number
9102383
Product problems
- Device-Device Incompatibility
- Device-Device Incompatibility
Patient
Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D4: LOT NUMBER ADDED. D4: EXPIRATION DATE ADDED. H3, H4, H6. PART : 03.231.007. LOT NO : 9102383. RELEASE TO WAREHOUSE DATE: 18 SEP 2014. MANUFACTURING SITE: WERK HAGENDORF. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM, EXHIBITS SIGNS OF NORMAL USE. NO SIGNIFICANT PRODUCT PROBLEM WAS OBSERVED ON THE SURFACE OF THE DEVICE. A DIMENSIONAL INSPECTION WAS PERFORMED FOR THE LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM AND MET SPECIFICATIONS. A FUNCTIONAL TEST TO ASSESS THE REPORTED ASSEMBLY ALLEGATION, WAS NOT CONDUCTED SINCE THE PLATE WAS NOT RETURNED FOR EVALUATION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM WAS FOUND TO HAVE NO DAMAGE OR DEFECTS. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DRAWING/SPECIFICATIONS REVIEWED: REV F CURRENT, D MANUFACTURED. DIMENSIONAL INSPECTION: DRAWING: D MANUFACTURED. FEATURE: SHAFT DIAMETER. SPECIFICATION: MEASURED DIMENSION: RESULT: CONFORMING MEASUREMENT DEVICE: CALIPER MITUTOYO CD. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. D9.
Description of Event or Problem
IT WAS REPORTED THAT ON (B)(6) 2022, THE LOCKING/NEUTRAL GUIDE FOR THE 4.5MM VA LCP CURVED CONDYLAR PLATE WAS MALFUNCTIONING AND WOULD NOT LOCK INTO PLACE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. NO PATIENT CONSEQUENCES WERE REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A LOCK/NEUTRAL GUIDE F/4.5MM VA LCP CRVD CONDYLAR AIMING ARM. THIS IS REPORT 4 OF 10 FOR (B)(4).
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: ADDITIONAL PROCODE: HWC, HRS. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER. REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.