OtherCFN

ADVIA 2400

Received Jan 21, 2010 · Event occurred Nov 23, 2009

Report 2432235-2010-00007 · MDR key 1585963

Device

Generic name

Chemistry Analyzer

Model number

ADVIA 2400

Catalog number

073-A010

Lot number

NA

Product problems

  • Low Test Results

Patient

63 YR

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

DISCORDANT IMMUNOGLOBULIN M_2 (IGM_2), RESULTS WERE OBTAINED ON AN ADVIA 2400 FOR 1 PATIENT. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE PHYSICIAN QUESTIONED THE RESULTS. DUE TO THE PATIENT'S HISTORY, THE PHYSICIAN EXPECTED A RESULT OF APPROXIMATELY 30 G/L. SUBSEQUENT ELECTROPHORESIS ON THE SAMPLE YIELDED A CONFIRMATORY RESULT OF 29 G/L. THERE WERE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES OR KNOWN PATIENT INTERVENTIONS DUE TO THE DISCORDANT IGM_2 RESULTS.

Additional Manufacturer Narrative

A SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE ENGINEER WAS NOT DISPATCHED TO THE CUSTOMER SITE. THIS APPEARS TO BE A CASE OF POSSIBLE SAMPLE INTERFERENCE. NO FURTHER CONCLUSION CAN BE DRAWN AT THIS TIME.

Remedial action

  • Other