inforMED
MalfunctionILZ

SAUNDERS CERVICAL TRCTN DVCE

Received Nov 17, 2022 · Event occurred Oct 21, 2022

Report 9616086-2022-00031 · MDR key 15820705

Device

Generic name

Accessories, Traction

Manufacturer

Djo Llc

Model number

199594

Product problems

  • Defective Component
  • Defective Component

Patient

Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT AFTER AIR PRESSURE IS APPLIED, ONCE RELEASED, THE AIR ALLEGEDLY COMES OUT IMMEDIATELY INSTEAD OF GRADUAL RELEASE. IT WAS REPORTED THAT PATIENTS ARE ALLEGEDLY GETTING HURT FROM QUICK RELEASE. THE DEVICE WAS RETURNED TO ENOVIS FOR EVALUATION. A FUNCTIONAL TEST COULD NOT BE DONE DUE TO THE PUMP BEING DAMAGED AND THE PRODUCT WILL NOT WORK CORRECTLY, THE MALFUNCTION REPORTED BY THE CUSTOMER HAS BEEN CONFIRMED. ENOVIS HAS REACHED OUT TO THE CUSTOMER FOR ADDITIONAL INFORMATION ON PATIENTS AND HAVE NOT RECEIVED ANY TO DATE.

Description of Event or Problem

IT WAS REPORTED THAT AFTER AIR PRESSURE IS APPLIED, ONCE RELEASED, THE AIR ALLEGEDLY COMES OUT IMMEDIATELY INSTEAD OF GRADUAL RELEASE. IT WAS REPORTED THAT PATIENTS ARE ALLEGEDLY GETTING HURT FROM QUICK RELEASE.