inforMED
InjuryHAM

CONMED

Received Dec 21, 2009 · Event occurred Oct 27, 2009

Report 1720159-2009-00048 · MDR key 1575149

Add to watch list
Product code (HAM)
Model (138030)

Device

Generic name

Argon Beam Coagulation, Nozzle

Catalog number

138030

Lot number

UNK

Product problems

  • Detachment Of Device Component
  • Loose or Intermittent Connection

Patient

Not reported

  • Perforation
  • Pneumothorax

Narrative

Description of Event or Problem

THE PT WAS UNDERGOING A LIVER TRANSPLANT. THE SURGEON WAS USING A CONMED ABC-ARGON BEAM COAGULATOR. THE TIPS ARE DISPOSABLE AND THE CORD IS GOOD FOR 100 USES ACCORDING TO THE MFR. IT APPEARS ON THIS OCCASION THAT THE METAL INSERT WHICH DOES NOT PROTRUDE BEYOND THE PT END HAD SOMEHOW BECOME LOOSE INSIDE THE CARRIER. WHEN THE GAS WAS APPLIED, THE METAL PIECE WAS SHOT OUT WITH SOME FORCE. THE METAL PIECE PUNCTURED THE PLEURAL SPACE THROUGH THE DIAPHRAM CAUSING A PNEUMOTHORAX AND REQUIRING A CHEST TUBE.

Additional Manufacturer Narrative

CONMED ELECTROSURGERY HAS MADE SEVERAL ATTEMPTS TO CONTACT THE INITIAL REPORTER. WE HAVE HAD NO REPLY TO OUR ATTEMPTS TO OBTAIN THE SUSPECT DEVICE FOR EVAL.