ADULT NASAL CANNULA WITH 2.1M OXYGEN TUBING
Received Oct 21, 2022 · Event occurred Jul 5, 2021
Report 3006061749-2022-00015 · MDR key 15650355
Device
Generic name
Nasal Cannula
Manufacturer
Flexicare Medical LtdCatalog number
032-10-022
Lot number
200303985
Product problems
- Appropriate Device Problem Term/Code Not Available
- Appropriate Device Problem Term/Code Not Available
Patient
Unknown
- Insufficient Information
- No Clinical Signs, Symptoms or Conditions
- Insufficient Information
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
EVENT WAS SUBMITTED TO THE (B)(6), THE ASSIGNED REFERENCE WAS (B)(4). DATE OF INITIAL SUBMISSION WAS 28 JULY 2021, FINAL SUBMISSION 01 SEPTEMBER 2021. THE DEVICE WAS RETURNED TO THE MANUFACTURER, WHERE SUBSEQUENT TESTING AND PRODUCT EVALUATION WAS CARRIED OUT. THE TESTING REVEALED NO IDENTIFIED DEFECTS WITH THE DEVICE, FURTHER COMMUNICATION WITH THE END USER HAS REVEALED THAT THE CANNULA HAS POTENTIALLY BEEN INCORRECTLY FITTED, WHERE THE SHOULDER HAS NOT BEEN SECURED UNDER THE PATIENTS NOSE TO KEEP THE DEVICE IN POSITION. OXYGEN SATURATION LEVELS ARE CONSTANTLY MONITORED, THEREFORE ANY DECREASE IN SUPPLEMENTARY OXYGEN WOULD BE EASILY NOTICABLE TO THE HEALTHCARE PROVIDER. THERE WERE NO REPORTED ADVERSE HEALTH EFFECTS ON THE PATIENT, THE ISSUE WAS IDENTIFIED BEFORE THE ALARMS NOTIFIED OF ANY DROP ON SATURATION LEVELS.
Description of Event or Problem
NASAL CATHETERS ROTATE AWAY FROM THE FACE, SO PT DOES NOT GET OXYGEN.